Technical Data 

Equipment type

Single Channel Digital TENS

Display

Digital LCD

Output Waveform

Bi-phasic rectangular

Auto Power - Off

5 Minutes

Programme Specifications

TENS 1

Freq. / Pulse Width : 80Hz / 125µs.

TENS 2

Freq. / Pulse Width : 2 Hz / 200µs.

TENS 3

Burst 2pps; 
Freq. / Pulse Width : 80 Hz / 150µs.

TENS 4

(Dense Disperse)
3 secs @ 80Hz with 100µs.
3 secs @ 2 HZ with 200µs.

Programme Freq.

P.W. 

On Time(s) Off Time(s)

Ramp 

Ramp 

(Hz)

(µs)

(Plus Ramp

On(s)

Off (s)

On)

EMS 1

50

125

4

3

1

N/A

EMS 2

50

150

5

5

1

N/A

EMS 3

30

200

N/A

N/A

3

3

Programme Output Frequency / Duration

PELV 1

5Hz / 3mins

10Hz / 5mins

15Hz / 5mins

20Hz / 5mins

15Hz / 5mins

10Hz / 5mins

5Hz / 2mins

Total time = 30mins

PELV 2

10Hz / 5mins

35Hz / 5mins

50Hz / 10mins

35Hz / 5mins

10Hz / 5mins

Total time = 30mins

PELV 3

MIXED

10Hz / 10mins

20Hz / 5mins

30Hz / 5mins

40Hz / 5mins

50Hz / 10mins

40Hz / 5mins

30Hz / 5mins

20Hz / 5mins

10Hz / 10mins

Total time = 60mins

17

Содержание Femme Elite

Страница 1: ...Femme Elite Instructions for use Pelvic Floor Stimulator TENS Pain Reliever and EMS Muscle Toner The 3 in 1 ...

Страница 2: ...you to live life to the full Before you use the Femme Elite it is important that you read this manual carefully it contains all the information you need to ensure you are using the unit safely and correctly Explanation of Symbols on Unit Equipment providing a particular degree of protection against electric shock partic ularly regarding allowable leakage currents having an F type floating applied ...

Страница 3: ... Incontinence the facts 8 When to use PFS 9 Using the Femme Elite for PFS 9 PFS FAQ s 10 TENS Pain Relief About TENS 11 What can TENS be used for 11 Using the TENS mode 12 Electrode pad placement 12 FAQ s and troubleshooting 13 EMS Muscle Toning About EMS 14 What can EMS be used for 14 Using the EMS mode 15 Electrode pad placement 15 EMS FAQ s 16 Technical data 17 Useful contacts 18 3 ...

Страница 4: ...ce such as a diaphragm or cervical cap If you have any concerns regarding use of the Femme Elite please do not hesitate to contact Body Clock Health Care on 020 8532 9595 General Precautions 1 Do NOT use this unit without first reading these instructions 2 Do NOT immerse the Femme Elite in any liquid 3 Do NOT place it close to any source of excessive heat 4 Do NOT operate this unit in the presence...

Страница 5: ...e or increase the intensity of the pulses down OFF key Press the down OFF key to either turn off the Femme Elite or decrease the intensity of the pulses In order to switch the Femme Elite off either turn the intensity down all the way until the unit switches off or press PROG key once and then the down OFF key Display symbols and their meanings Mode indicator Indicates mode in use Programme indica...

Страница 6: ...upward Replace the cover see fig 1 page 7 b Connect the leadwire adaptor to the Femme Elite unit by inserting the protruding pin on the adaptor into the holes on the back of the Femme Elite Slide upwards to ensure it is fixed securely see fig 2 page 7 When using the Pelvic Floor Stimulation mode Connect the Elite vaginal probe according to the instructions found in probe packaging Switch the Mode ...

Страница 7: ...7 Useful Diagrams ...

Страница 8: ...ce occurs when sneezing coughing laughing lifting or during exercise and causes a small amount of urine to leak 2 Urge incontinence occurs with a strong sudden and uncontrollable urge to urinate 3 Mixed incontinence a combination of urge and stress incontinence The Femme Elite is designed to stimulate and strengthen your pelvic floor muscles for you It has 3 PFS programmes each treats a different ...

Страница 9: ...robe packaging c Switch the unit on d Select the Programme designed for your type of incontinence by using the PROG key e Gradually increase the intensity until you can comfortably feel the muscles contacting f At the end of the programme the Femme Elite will automatically switch off g After you have finished remove and thoroughly clean the vaginal probe in accordance with the vaginal probe instru...

Страница 10: ...type of incontinence from which you are suffering STRESS Incontinence improvement comes slowly with the building of muscle fibre and nerve sensitivity This can take up to 2 months URGE Incontinence improvements are made by reducing the false signals sent by the bladder to the brain Results can be seen in as little as 14 days MIXED Incontinence improvements are made by building the muscle fibre and...

Страница 11: ...ite you need to set up the unit see page 6 Ensure that the mode selector is switched to TENS before commencing treatment The Femme Elite has 4 TENS programmes each treats a different type of pain TENS 1 Back Pain TENS 2 Migraine TENS 3 Period Pain TENS 4 Shoulder Pain If you prefer you can use any of the TENS programmes to treat your pain The above table shows which pain areas the programmes are c...

Страница 12: ...n the screen If it does not switch the unit off and adjust the mode selector to TENS and switch the unit on again You can then choose your programme using the PROG key Once you have selected your programme gradually increase the intensity using the ON Up key until the pulsing sensation is comfortable TENS Programme 1 Back Pain TENS Programme 2 Migraine TENS Programme 3 Period Pain TENS Programme 4...

Страница 13: ... Care on 020 8532 9595 Q Can I use this TENS for labour A It is not recommended It does not feature the boosterbuttonrequiredforuseduringcontractions For information on TENS for use during labour please visit www BabycareTENS com or call 0845 230 9737 Troubleshooting Questions Q Why does the pulse sensation not appear to be as strong after you have used the Femme Elite for a while A Increase the i...

Страница 14: ... and build the abdominal muscles It has 3 different EMS modes each is a specifically designed exercise programme for the muscle groups women tend to want to treat e g abdominal and thigh muscles If this is the first time you are using your Femme Elite you need to set up the unit see page 6 Ensure that the mode selector is switched to EMS before commencing treatment The Femme Elite has 3 EMS progra...

Страница 15: ...switch the unit off and adjust the mode selector to EMS Then switch the unit on again You can then choose your programme using the PROG key Select the programme suitable for the muscle group you wish to tone Gradually increase the intensity using the ON Up key until the tensing sensation is comfortable The intensity level appears as blocks on the bottom of the screen see page 5 EMS Programme 1 Abd...

Страница 16: ... to woman Some women may see results very quickly for others it may take several weeks Q Can I use it after giving birth A You can do but not until about 12 weeks after the birth Consult your health visitor if you wish to use it before then Q What is the difference between the EMS on the Femme Elite and the muscles toners you can buy on the high street A Technologically there is much similarity Ho...

Страница 17: ...s 3 secs 2 HZ with 200µs Programme Freq P W On Time s Off Time s Ramp Ramp Hz µs Plus Ramp On s Off s On EMS 1 50 125 4 3 1 N A EMS 2 50 150 5 5 1 N A EMS 3 30 200 N A N A 3 3 Programme Output Frequency Duration PELV 1 5Hz 3mins 10Hz 5mins 15Hz 5mins 20Hz 5mins 15Hz 5mins 10Hz 5mins 5Hz 2mins Total time 30mins PELV 2 10Hz 5mins 35Hz 5mins 50Hz 10mins 35Hz 5mins 10Hz 5mins Total time 30mins PELV 3 ...

Страница 18: ...o uk 0845 230 9737 The Continence Foundation Confidential Information and advice about bladder problems from specialist nurses www continence foundation org uk 0845 345 0165 Promocon Independent and impartial information on continence products and services www promocon2001 co uk 0161 834 2001 Incontact National charity for people affected by bladder and bowel problems www incontact org 0870 770 32...

Страница 19: ...19 ...

Страница 20: ...td 108 George Lane South Woodford London E18 1AD Tel 44 0 20 8532 9595 Fax 44 0 20 8532 9551 www bodyclock co uk e mail customercare bodyclock co uk 0120 January 2010 ISO 9001 2008 ISO 13485 2003 EC DIRECTIVE 93 42 EEC Annex V ...

Отзывы: