background image

2.4.4 Recovery of liquid in sample cell to Vessel 1 

 
The liquid in the sample cell cannot be recovered to Vessel 1 using the liquid 
handling rig as presently configured. 
 

2.4.5 Recovery of liquid in sample cell to Vessel 2 

 
The liquid in the sample cell can be recovered to Vessel 2 using the following 
procedure. 
 

• 

Vessel 2 should be under vacuum along with the manifold between valves V5 
and V6 prior to the recovery of the liquid from the sample cell.  If it is not, 
switch the pump to roughing mode and open tap E and valve V7 followed by 
valves V1 and V2.  If Vessel 2 already contains some liquid, ensure that it is 
frozen using liquid nitrogen before evacuation is attempted. 

• 

When the vessel and manifold have been evacuated, close tap E and cool the 
glass bulb by raising a dewar of liquid nitrogen around it. 

• 

Ensure that tap E and valves V6 and V8 are closed before attempting to open 
valve V5. 

• 

Opening valve V5 will allow the liquid present in the sample cell to be drawn 
into Vessel 2 where it will freeze. 

• 

All the liquid may not return in a single pass and repeating the evacuation 
procedure may be required.  The liquid may also be encouraged to leave the 
sample cell by using a small amount of gas. 

• 

Close valve V3 and open valve V9 to allow helium gas into this part of the 
system.  Close valve V9 and carefully open valve V4 to allow the gas to enter 
the sample cell.  DO NOT repeat this procedure until the pressure in the 
system is known by opening tap E. 

• 

Close valve V7 and re-pressurise the system and Vessel 2 by opening valve 
V9 and tap E. 

 

2.4.6 Recovery of liquid in sample cell to Vessel 3 

 
The liquid in the sample cell can be recovered to Vessel 3 using the following 
procedure. 
 

• 

Vessel 3 should be under vacuum prior to the recovery of the liquid from the 
sample cell.  If it is not, switch the pump to roughing mode and open tap C 
and valve V3 followed by valves V1 and V2.  If Vessel 3 already contains 
some liquid, ensure that it is frozen using liquid nitrogen before evacuation is 
attempted. 

• 

When the vessel and manifold have been evacuated, close valve V3 and cool 
the glass bulb by raising a dewar of liquid nitrogen around it. 

• 

Ensure that tap D and valves V6, V7 and V8 are closed before attempting to 
open valve V4. 

• 

Opening valve V4 will allow the liquid present in the sample cell to be drawn 
into Vessel 3 where it will freeze. 

 - 

11 

Содержание Muon Liquid Handling System None

Страница 1: ...Muon Liquid Handling System User Guide C Johnson S P Cottrell et al Version 0 2 1...

Страница 2: ...essel 2 7 2 2 3 Removing air from Vessel 3 7 2 2 4 Evacuation of the sample cell 7 2 3 Freeze Pump Thaw cycle 8 2 3 1 Sample in Vessel 1 8 2 3 2 Sample in Vessel 2 8 2 3 3 Sample in Vessel 3 9 2 4 Liq...

Страница 3: ...uid handling system are shown in Figure 1 Inlet 1 Inlet 2 Gas pressure Cell pressure To pump To cell Baratron Vessel 2 Vessel 1 Vessel 3 Figure 1 The layout of the front panel on the muon liquid handl...

Страница 4: ...e a means of flowing liquid into and out of the cell along with feed throughs for an RF excitation signal and a pick up coil Figure 2 The liquid sample cell mounted on a sample stick with a flat coil...

Страница 5: ...Roughing Turbo switch Off Run switch Pressure reading Turbo control panel Turbo Pump Rotary Pump Figure 2 The layout of the front panel on the vacuum pump used with the muon liquid handling system 5...

Страница 6: ...ls It is important to note that a small amount of liquid may remain in the system after sample loading is finished Care must be taken to prevent this liquid reaching the pump 2 2 Evacuation of the sys...

Страница 7: ...n the sample is frozen a vacuum can be achieved by switching the pump to its roughing mode and opening valve V1 For a high vacuum switch the pump to turbo 2 2 3 Removing air from Vessel 3 Once the sys...

Страница 8: ...be read from the baratron gauge Tap B should be closed to prevent suck back of the liquid as the solid thaws The sample can be warmed gently to aid the thawing process and when complete close tap A 2...

Страница 9: ...t by cooling evacuation or both 2 4 1 Vessel 1 to sample cell The transfer of liquid from Vessel 1 to the sample cell is brought about by the difference in pressure between that in the Vessel and a va...

Страница 10: ...the sample cell When it is judged that sufficient liquid has been transferred to fill the sample cell close valve V5 At this point it must be remembered that the lower manifold contains liquid which m...

Страница 11: ...e evacuation procedure may be required The liquid may also be encouraged to leave the sample cell by using a small amount of gas Close valve V3 and open valve V9 to allow helium gas into this part of...

Страница 12: ...liquid may also be encouraged to leave the sample cell by using a small amount of gas Close valve V3 and open valve V10 to allow helium gas into the lower part of the system Close valve V10 and carefu...

Страница 13: ...he rig Muon spin resonance experiments were performed at 2 G and 20 G followed by reattached to the rig At this point the rig was open to the air and so had to be evacuated and pumped for 30 minutes T...

Страница 14: ...ond experiment all of the n hexane used in the first experiment was recovered into a sample bottle and dried using a small piece of clean sodium The liquid was carefully poured back into Vessel 1 and...

Отзывы: