Operation
EASY- CHECK FN
Manual
Rev:00
5 / 10
<Prn> Print or activate the RADIO CONTROL function
When EASY-CHECK FN is connected to the printer MEGA-PRINT or operated with the
RADIO CONTROL module, press the key until
<Prn>
is displayed and wait until
<Prn>
is
flashing. Start the data transfer by pressing the key.
This is not necessary when working with the interface cable and the software
selected. As soon as the interface cable is connected, the data transfer starts by
itself without pressing the key.
<IcL> Clear the last reading
Press the key until the symbol
<IcL>
is displayed and wait until
<IcL>
is flashing. Confirm
by pressing the key again.
Now the last reading is cleared.
<cLr> Clear the whole Data Memory
Press the key until the symbol
<cLr>
is displayed and wait until
<cLr>
is flashing. Confirm
by pressing the key again.
Now the whole Data Memory is cleared.
<nFE> Base material recognition
This mode is necessary when you want to measure paint or other insulating coatings on
zinc plated steel.
In the Standard mode (FE + nFE) the device would measure the paint and the zinc coating
together, because it is not able to recognize the thin zinc coating as non-ferrous base ma-
terial nFE.
To measure only the paint coating on the zinc plating the device must be switched over to
the nFE mode.
To avoid wrong measurements the following requirements should be met:
•
The zinc plating should have a thickness of at least 20 µm.
•
The zero adjustment should be performed on a similar zinc plated object without
coating.
The nFE Mode should also be used when measuring on slightly magnetic stainless steel.
From factory the device is adjusted to the FE + nFE recognition mode. To switch over to
the nFE mode press the key until the symbol
<nFE>
is displayed. As soon as
<NFE on>
is
flashing confirm by pressing the key again.
As soon as the device is switched off the nFE recognition is deactivated and the
device will measure again in the FE + nFE mode.
<Un> Change unit of measurement (µm / mils)
In its basic setting the instrument measures in µm.
To measure in “mils” (American unit of measurement) press the key until the symbol
<un>
is displayed. Release the key and wait until
<on µm>
is flashing. Press the key again to
switch over to
<on mils>
.
When you switch on again the device it measures in “mils”.
To change over to “µm” proceed in the same way.