BMS C1
II. Mechanical Part:
PROBLEM
REASON FOR PROBLEM
SOLUTION
1.
Can’t focus when use
high magnification
objective.
Slice put reversed.
Cover slice is too thick.
Reverse back
Using standard cover slice.
Thickness: 0.17 mm
2.
Touch the slice
when switch the
objective from low
magnification to
high one.
Slice put reversed.
Cover slice is too thick.
Reverse back
Using standard cover slice.
Thickness: 0.17 mm
3.
The specimen move
not flowing
The slice holder not nip the
slice firmly.
Make it firmly
4.
Two eyes image not
in superposition
The interpupillar distance is not
correct
Adjust
the
interpupillar
distance correctly
The diopter is not right
Adjust the diopter according
your sight
5.
The eyes are
uncomfortable
6.
The bright degree is not
properly
Adjust the voltage of bulb.
III. Electric Part:
PROBLEM
REASON FOR PROBLEM
SOLUTION
No power supply
Check the power cord, and
connect them exactly
the installation of the bulb is wrong
Install the bulb correctly
1.
The lamp can’t
light
The bulb burn out
Change a new bulb
2.
The bulb burn out
in a high frequency
Not use a appointed lamp
High voltage
Use a appointed lamp. If the
circumstance still not change,
please contact the maintenance
part.
3.
The rightness degree
is not enough
Not use a appointed lamp
Low voltage
use a appointed lamp
Turn up the voltage
The bulb is going to spoil
Change the bulb
4.
The light glimpse
The power cord have a poor
contact
Check the power cord, and
connect them exactly