3.3.4 Adjusting the Condenser(Fig.24) 

Turn the condenser focus knob to shift the condenser. It 

needs to raise the condenser when using the high 

magnification objective, and to decline when using the 

low magnification one. 

 



 The center of the condenser and the light axes 

of  the  objective  are  coaxial.  It  has  been 
adjusted before leaving factory, so the user 
needn’t to adjust them by self. 



 The highest position of the condenser has been 

adjusted  too.  It  also  needn’t  any  user’s 
operation. 

3.3.5 Adjusting the Aperture Diaphragm 
(Fig.25)

 

Turn the

 

Aperture Diaphragm Lever①,to adjust 

the aperture size. 



  Aperture Diaphragm is designed for adjusting 

the  aperture  size,  not  for  adjusting 
brightness. 

Generally, reducing the diaphragm opening to 
70- 80% of the N.A. value of the respective 
objective will provide an image of acceptable 
quality. If you want to observe the image of 
the aperture diaphragm, remove one eyepiece 
and look through the tube. You will see a dark 
circle encroaching on the bottom of the tube. 

 

3.3.6 Adjusting the Interpupillar Distance 
(Fig.26) 

     The interpupillar distance range:

55mm~75mm

When observing with two eyes, hold on the left and 

right prism holder, turn around the axis, adjust the 

interpupillar distance until the left and right fields 

of view coincide completely.

 

10 

Fig.24 

Fig.25 

Fig.26 

Содержание BMS C1 series

Страница 1: ...his manual is written for biological microscope BMS C1 For safety exerting best performance of the instrument and making you familiar with the instrument entirely we strongly recommended that you carefully read this manual before using the microscope ...

Страница 2: ...onents 1 2 Installation 2 2 1 Installing Illustration 2 2 2 Installing Steps 3 3 Adjustment Operation 7 3 1 Adjustment System Installing Illustration 7 3 2 Operation 9 4 Technical Specifications 12 5 Configuration Table 13 6 Trouble Shooting 13 ...

Страница 3: ...mpletely cool 8 Check the input voltage be sure the input voltage which signed in the back of the microscope is consistent with the power supply voltage or it will bring a serious damage to the instrument 9 Use the factory supplied power cord please II Maintenance 1 All the lenses have been well checked and adjusted It is forbidden to disassemble them yourself 2 The nosepiece and coarse fine focus...

Страница 4: ...pe That will affect the function or decline the performance of the microscope 8 Place the instrument in a cool dry position After using the microscope remember to cover it with dust helmet Do wait for the lamp house cooling completely before cover ...

Страница 5: ...1 NAME OF SPARE PARTS BMS C1 bino objective Eyepiece Binocular viewing head Nosepiece Objective Body 1 ...

Страница 6: ...r express the installing process before the installing make sure there is no dust and stain on the spare parts No marks by outside force on the surface of the spare parts or glass Keeping the hexagon spanner well when change the spare parts you will use it Body Objective Binocular Viewing Head 10x Objective 2 Cable Wire ...

Страница 7: ...viewing head picture No 1 2 Insert the binocular viewing head in the head of the body turn to the right place then fix up it by bolts 2 2 2 Install the eyepiece picture No 3 4 Insert the eyepiece in the eyepiece tube till end No 4 picture shows the station after the installing Picture 4 ...

Страница 8: ...2 Screw down the objective to nosepiece one by one Clean the objective frequently At first use the 10X objective to looking for image then change another one Turn the objective till hear the kaca sound make sure the objective enter the objective center 2 2 4 Install the filter picture 7 1 Swing out the condenser holder 2 Put the filter in the holder Then swing the holder in ...

Страница 9: ...nsert the electronic plug in the socket safely 3 Insert on Use the cable wire supplied by BMS company when destroyed make the choice of same type The machine adopt the wide voltage can work under the voltage of 230V 50Hz 2 2 6 change the fuse picture9 10 Before change the fuse make sure that the switch is cut off swing out the change new fuse then swing back into the 5 Picture 8 Picture 9 Picture ...

Страница 10: ...Pull out the condenser set 1 Use the pledge wrap the lamp put the lamp pin into the socket 2 Change the lamp during usage Make sure the switch is cut off all parts around lamp are cool down Insert the lamp carefully do not touch the lamp with finger strong force Insert the condenser opposite direction as picture 6 Picture 13 ...

Страница 11: ... adjustment system installing illustration picture No 14 15 left fine focus knob left coarse focus knob 7 interpupillary distance showing Eye viewing adjust ring light adjusting switch Picture 14 locking knob condenser adjustment knob ...

Страница 12: ...BMS C1 Tension Adjustment Collar Right Fine Focus Knob Main Switch Portrait Adjustment Knob Lateral Adjustment Knob 8 15 Right Coarse Focus Knob Aperture Diaphragm Lever ...

Страница 13: ...ective If you finish the observation with the short working distance objective and want to change another one be careful of not letting the objective touch the specimen 3 3 3 Focusing Fig 23 1 Use the 10 objective focus to avoid the objective touch with the specimen you should raise the mechanical stage at first let the specimen close to the objective then slowly separating them to focus The opera...

Страница 14: ...re Diaphragm Lever to adjust the aperture size Aperture Diaphragm is designed for adjusting the aperture size not for adjusting brightness Generally reducing the diaphragm opening to 70 80 of the N A value of the respective objective will provide an image of acceptable quality If you want to observe the image of the aperture diaphragm remove one eyepiece and look through the tube You will see a da...

Страница 15: ...sion adjustment collar has adjusted before leaving factory if finding it s loosing the mechanical stage drop itself because of deadweight please turning the tension adjustment collar until the tightness is in order Turn it along the direction show in the picture the Coarse Focus Knob will become tighter Turn anti direction will become loosen If the mechanical stage drop itself or even lose focus j...

Страница 16: ...ser Abel NA 1 25 Stage Double layer mechanical stage area 140 140mm movement range 75 50mm Lamp House halogen lamp 6V 20W Eyepiece Objective 1 Objective Magnification Numerical Value Aperture Diaphragm N A Thickness of Cover Slip mm Focus mm Working Distance mm Working Mode 4 0 10 0 17 31 05 37 5 Dry 10 0 25 0 17 17 13 7 316 Dry 40 0 65 0 17 4 65 0 632 Dry 100 1 25 0 17 2 906 0 198 Oil 2 Eyepiece ...

Страница 17: ...Head 1 〇 Condenser N 1 25 Abbe Condenser 1 〇 Nosepiece Quadruple Nosepiece 1 〇 6V 20W Halogen Lamp 1 〇 Spare Lamp 2 〇 Lamp house Spare Fuse BGX 1 20 0 5A 2 〇 2 Eyepiece Plan Eyepiece 10 2 〇 Achromatic Objective 4 1 〇 Achromatic Objective 10 1 〇 Achromatic Objective 40 1 〇 Objective Achromatic Objective 100 Oil Spring 1 〇 Condenser Bright Field Condenser with Changeable Raster 1 〇 Filter Baby Blue ...

Страница 18: ...pecimen is reversed Reverse back The dry objective is stained with oil Especially 40X Wipe up Stains on the lens Condenser Objective Eyepiece Wipe up The oil objective is not in oil Use oil There is air bubble in oil Remove oil Not use the appointed oil Use the appointed oil The opening of Aperture diaphragm is too large Properly make it smaller Stains on the incidence lens in the binocular drawtu...

Страница 19: ...ion Turn it into the required position 6 The image takes the yellow slightly Not use the color filter Use the blue filter The opening of Aperture diaphragm is too small Properly make it larger Condenser position is too low Adjust the position 7 The bright degree is not enough Stains on the lens Condenser Objective Eyepiece Wipe up 14 ...

Страница 20: ...ter is not right Adjust the diopter according your sight 5 The eyes are uncomfortable 6 The bright degree is not properly Adjust the voltage of bulb III Electric Part PROBLEM REASON FOR PROBLEM SOLUTION No power supply Check the power cord and connect them exactly the installation of the bulb is wrong Install the bulb correctly 1 The lamp can t light The bulb burn out Change a new bulb 2 The bulb ...

Отзывы: