Page 8

2.3   Input Power Connections

 

 

Risk of electric shock – cord connected models are supplied with a grounding 
conductor and grounding-type attachment plug. To reduce risk of electric shock, be 
certain that it is connected only to a properly grounded, grounding-type receptacle.

Three power cord plug types available. 

Power cord length is 6 feet (3.83 meters)

WARNING

!

Be certain to connect pump to proper supply voltage.  Using incorrect voltage will damage pump and may result in 

injury.  Voltage requirement is printed on pump serial label.

!

Input power range is 100VAC to 240VAC 50/60 Hz.

!

Voltage Selection is automatically detected and adjusted by power supply. No mechanical switch necessary.

!

Use voltage your power cord is rated for.

!

Cord connected models are supplied with a ground wire conductor and a grounding type attachment plug (power 

cord). To reduce risk of electric shock, be certain that power cord is connected only to a properly grounded, 
grounding type receptacle.

!

Permanently connected models must be properly grounded. Be certain that a grounding conductor is connected 

to terminal T11-1 located in the wiring compartment.

!

Wiring compartment access requires removing 4 screws. A 5/32” (.156”) allen wrench is required (included).

!

Be sure all wiring compartment cable glands and hole plugs are properly installed and sealed.  

!

Never strap control (input / output) cables and power cables together.

!

Power Interruption:

 This pump has a user programmable auto-restart feature which will either restore the pump 

to the operating state it was in when power was lost or require a user action to restart.

Note:

  When in doubt regarding your electrical installation, contact a licensed electrician.

Electrical connections and grounding (earthing) must conform to local wiring codes. 
Be certain that a grounding conductor is connected to terminal T11-1 located in the 
wiring compartment.

WARNING

115V 60Hz
NEMA 5/15 (USA)
max: 125V AC

230V 60Hz
NEMA 6/15 (USA)
max: 250V AC

240V 50Hz
CEE 7/VII (EU)
max: 250V AC

QTY. DESCRIPTION

2

.50 INCH (12.7 mm) LIQ-TIGHT HOLE PLUGS (MAT’L = NEOPRENE), PRE-INSTALLED

3

.875 INCH (22.2 mm) LIQ-TIGHT HOLE PLUGS (MAT’L = NEOPRENE), 2 PRE-INSTALLED

2

.50 INCH (12.7 mm) LIQ-TIGHT CONNECTORS FOR PASS THRU CORDS (MAT’L = NYLON)
ACCEPTABLE CABLE DIAMETER .118 TO .255 INCH (3.0 TO 6.5 MM), NOT INSTALLED

3

.875 INCH (22.2 mm) METALLIC LIQ-TIGHT CONNECTORS FOR PASS THRU CORDS (MAT’L = NYLON)
ACCEPTABLE CABLE DIAMETER .200 TO .395 INCH (5.1 TO =10.0 MM), 1 PRE-INSTALLED WITH POWER CORD MODELS

Cable and conduit connectors included

5.99  in.

(152.1  mm)

Ø  .50
(12.7  mm)
2-PLCS.

Ø  .84  in.
(21.3  mm)
3-PLCS.

WIRING COMPARTMENT COVER

POWER CORD OPTIONS

Risk of electric shock - Disconnect electricity before removing the wiring 
compartment cover.

WARNING

MD-3

2

METALLIC LIQ-TIGHT CONNECTORS FOR .50 INCH FLEXIBLE CONDUIT (MAT’L = DIE CAST ZINC), NOT INSTALLED

Page 9

OUTPUT

INPUT

3

1
2

T3

DRY

RMT

COM

1
2
3

T1

PULSE

GND  (-)

4-20  (+)

0-10  (+)

+15 VDC

1
2

T2

T4

3

1 2

(+)

FVS

SIG

(-)

2

1

FREQ.

(-)

(+)

T5

1 2

4-20 

mA

T6

POWER

FUSE  5A  SLOW  BLOW
(20  X  5MM)

F1

2

3

1

GROUND

N EUTRAL
(COMMON)

LINE
(HOT)

T11

N.C.

COM

N.O.

N.O.

COM

N.C.

CC3

CC2

T8

T9

3

1
2

3

1
2

3

1
2

CC1

N.O.

COM

N.C.

T10

1 2 3

N.C. COM.

N.O.

RELAY

 OUT

T7

(-)

(+)

2.4   Wiring Terminals and I/O Schematics

 

 

INPUT:
4-20 mA

INPUT:
0-10V DC

INPUT:
FREQUENCY, AC
SINE WAVE, TTL,
CMOS

FUNCTION

TERMINAL PIN #

RATING

ELECTRICAL SP.

BLOCK DIAGRAM

T1

T1

T1

T1

T2

T1

INPUT:
FVS SYSTEM 
(FLOW VERIFICATION
SENSOR)
FV SENSOR ONLY

T4

T4

T4

(+) POSITIVE

3

2

1

2

1

2

1

2

3

(-) NEGATIVE

(+) POSITIVE

(-) NEGATIVE

(+) POSITIVE

(-) NEGATIVE

(+) POSITIVE

SIGNAL

(-) NEGATIVE

INPUT:
REMOTE
START / STOP
(DRY CONTACT C.)

INPUT:
REMOTE
START / STOP
(WET CONTACT C.)

OUTPUT:
4-20 mA

OUTPUT:
FREQUENCY - 
OPEN COLLECTOR

OUTPUT:
RELAY, 6 AMP

OUTPUT:
CONTACT
CLOSURE 1

INPUT:
POWER

OUTPUT:
CONTACT
CLOSURE 2

OUTPUT:
CONTACT
CLOSURE 3

FUSE

T3

T3

1

2

T3

T3

2

3

T6

T6

2

1

T5

T5

2

1

(+) POSITIVE

(-) NEGATIVE

(+) POSITIVE

(-) NEGATIVE

T7

1

T7

T7

2

3

NORM. CLOSED

COMMON

NORM. OPEN

T8

1

T8

T8

2

3

NORM. OPEN

COMMON

NORM. CLOSED

T9

1

T9

T9

2

3

NORM. OPEN

COMMON

NORM. CLOSED

T10

1

T10

T10

2

3

NORM. OPEN

COMMON

NORM. CLOSED

T11

1

T11

T11

2

3

GROUND

NEUTRAL

LINE (HOT)

F1

NA

5 AMP

120 OHM
IMPEDANCE, NON
POWERED LOOP

3 TO 30 VOLT DC
1 AMP MAX.

120 OHM
RESISTANCE
ACTIVE LOOP

OPEN COLLECTOR
0-1000 Hz
50% DUTY CYCLE

Form C
6 AMP MAX AT
250 VAC,
5 AMP MAX AT
30 VOLT DC

Form C
1 AMP MAX AT
125 VOLT AC,
0.8 AMP MAX AT
30 VOLT DC

Form C
1 AMP MAX AT
125 VOLT AC,
0.8 AMP MAX AT
30 VOLT DC

Form C
1 AMP MAX AT
125 VOLT AC,
0.8 AMP MAX AT
30 VOLT DC

5A SLOW BLOW
(20 X 5MM)

1
2
3

T1

PULSE

GND  (-)

4-20  (+)

1
2
3

T1

PULSE

GND  (-)

4-20  (+)

0-10  (+)

+15 VDC

1
2

T2

0-1000 HZ MAX.

(-) NEGATIVE

(+) POSITIVE

(+) POSITIVE

(-) NEGATIVE

NO VOLTAGE

96 TO 264 VOLT AC,
50 / 60 HZ
A3 = 220W
A4 = 350W

1
2
3

T1

PULSE

GND  (-)

4-20  (+)

(+)

(-)

(+)

(-)

(+)

(-)

T4

3

1 2

(+)

FVS

SIG

(-)

RED  (+)

BARE

BLACK  (-)

3

1
2

T3

DRY

RMT

COM

3

1
2

T3

DRY

RMT

COM

(+)

(-)

(+)

(-)

2

1

FREQ.

(-)

(+)

T5

DIGITAL PULSE
RECEIVER CIRCUIT

1.5K OHM

(+)

(-)

EXTERNAL SOURCE
6-30V DC

1 2 3

N.C. COM. N.O.

RELAY

 OUT

T7

NO

C

NC

SWITCH LOAD
6 AMP MAX @ 250V AC
5 AMP MAX @ 30V DC

2

3

1

GROUND

N EUTRAL

(COMMON)

LINE

(HOT)

T11

N.C.

COM

N.O.

T8

3

1
2

CC1

N.O.

COM

N.C.

CC2

T9

3

1
2

CC3

3

1
2

N.O.

COM

N.C.

T10

NO

C

NC

SWITCH LOAD
1 AMP MAX @ 125V AC
0.8 AMP MAX @ 30V DC

NO

C

NC

C

NC

NO

SWITCH LOAD
1 AMP MAX @ 125V AC
0.8 AMP MAX @ 30V DC

SWITCH LOAD
1 AMP MAX @ 125V AC
0.8 AMP MAX @ 30V DC

EXTERNAL DEVICE
3 TO 30V DC

OPEN CIRCUIT
IMPEDANCE MUST
BE GREATER THAN
50K OHM

BLUE-WHITE
FVS SENSOR

0-10V DC
TRANSMITTER
SOURCE

FREQUENCY
TRANSMITTER
SOURCE

ACTIVE 4-20mA
TRANSMITTER
SOURCE

2

1

4-20mA

(-)

(+)

T6

4-20mA RECEIVER
600 OHM LOAD MAX.

(+)

(-)

96 TO 264 VOLT AC,
50 / 60 HZ
A3 = 220W
A4 = 350W

Risk of electric shock - All 
wiring must be  insulated 
and rated 300V minimum.

WARNING

MD-3

Terminals T1 thru T10
Plug type
16 - 24 AWG

Power Input Terminal T11
Plug type
14 - 30 AWG

T4

T4

T4

1

2

3

(+) POSITIVE

SIGNAL

(-) NEGATIVE

T4

3

1 2

(+)

FVS

SIG

(-)

SIGNAL

NEGATIVE  (-)

BLUE-WHITE
MICRO-FLO
FLOWMETER
PULSE OUTPUT

INPUT:
FVS SYSTEM 
(FLOW VERIFICATION
SENSOR)
FS or FP MICRO-FLO
FLOWMETER ONLY

1
2
3

T1

PULSE

GND  (-)

4-20  (+)

(OPTIONAL PUMP #2)

Single or dual pump (series) 
input. Loop voltage must not 
exceed 24 Volts.

Shielded cables should be used on all input signal wires.

NOTE:

 USE ONLY

DRY CONTACT FOR
REMOTE S/S WHEN 
USING 4-20mA 
INPUT OR 0-10V DC

Содержание PROSERIES-M CHEM-PRO MD-3

Страница 1: ...CA 92649 USA Phone 714 893 8529 FAX 714 894 9492 E mail sales blue white com or techsupport blue white com URL www ProSeries M com PROSERIES M MD 3 Operating and Maintenance Manual ProSeries by Blue...

Страница 2: ...Input Hz 19 4 3 5 Pulse Batch low speed batch 20 4 3 6 Manual Cycle Adjust repeating cycle timer 21 4 3 7 Dispensing 22 4 3 8 Manual Dosing 23 4 3 9 Proportional Dosing 24 4 4 Contact Closure Input Re...

Страница 3: ...w Adapters 1 2 MPT Footvalve 9 230V 50HZ power cord BS 1363 A plug United Kingdom MD 3 Multi Diaphragm Metering Pump 2 190 RPM 4 090 stroke large diaphragm X Standard Equipment includes power cord and...

Страница 4: ...O O O O 14 F to 115 F 10 C to 46 C Ambient Storage Temperature O O O O 40 F to 158 F 40 C to 70 C Operating Voltage 96 to 264VAC 50 60Hz 220 Watts Power Cord Options 115V60Hz NEMA 5 15 USA 230V60Hz N...

Страница 5: ...dard Compatible with Blue White s output Flow Verification Sensor FVS system Includes stainless steel extended mounting brackets Lifts pump 4 1 2 11 43 cm for easy access in hard to reach areas MD 3 E...

Страница 6: ...w prevention check valve at discharge side of pump to prevent system fluid from flowing back through pump during pump maintenance A pressure relief valve is recommended at discharge of pump Bolt pump...

Страница 7: ...22cm 8 86in 22 50cm 4 00in 10 15cm n Made out of tough Stainless Steel n Provides a stable mounting surface Stainless Steel extended brackets allow the pump to be securely mounted to most any surface...

Страница 8: ...ly installed and sealed Never strap control input output cables and power cables together Power Interruption This pump has a user programmable auto restart feature which will either restore the pump t...

Страница 9: ...DC Form C 1 AMP MAX AT 125 VOLT AC 0 8 AMP MAX AT 30 VOLT DC 5A SLOW BLOW 20 X 5MM 1 2 3 T1 PULSE GND 4 20 1 2 3 T1 PULSE GND 4 20 0 10 15VDC 1 2 T2 0 1000 HZ MAX NEGATIVE POSITIVE POSITIVE NEGATIVE N...

Страница 10: ...values by one and increase pump output in Manual Speed Adjust mode Press and hold to increase values progressively faster Press and release to stop the pump Press and release to select menu items Pres...

Страница 11: ...l bar example While making a selection where only one choice is allowed you will see a radio button Radio button example In a screen where you are making changes you will see the word Done located at...

Страница 12: ...en to exit the menu structure and enter the run mode MD 3 Configuration Display Language Set Password Display Rate Esc Reset to Default Display Language English Spanish German Done French Display Lang...

Страница 13: ...y may be toggled through Speed RPM and your selected Display Rate by pressing the right arrow button Press MENU button to enter the menu structure Select Configuration and Press ENTER button Select Di...

Страница 14: ...ate Input Max Flowrate Flowrate ML min Press ENTER to set 0000 0 1 Contact Input Close Stop Pump Open Stop Pump Input Time Delay OFF 000 Range 1 255 sec Disable 00 0 Input Setup 4 20 mA Input Frequenc...

Страница 15: ...off Max Flowrate Input Modes Contact Input Set FVS Esc Hidden Remote Local Set DFD Set Rev Alarm Max Flowrate Input Max Flowrate Flowrate ML min Press ENTER to set 0000 0 1 To calculate the Max Flowra...

Страница 16: ...ate back out of the menu structure To do this you must select Esc at the bottom of every screen menu until you see the Run Mode screen displayed Tip The Manual Speed Adjust mode can be combined with C...

Страница 17: ...ontinue this process until all four screens have been configured If you used short cut to enter 4 20 mA input setup press and hold the Mode button until the Run Mode screen is displayed If you used Me...

Страница 18: ...displayed in the top line of the display To configure the pump navigate to 0 10 VDC Input menu by using the short cut method described at the beginning of the section or by pressing MENU button then...

Страница 19: ...GHT ARROW to select the digit to change press UP or DOWN arrow to scroll through 0 9 on selected digit Press enter to save changes and move to the next input screen To input flow rate values press the...

Страница 20: ...easily scroll back by continuously pressing the RIGHT button Press ENTER to save the changes Continue this process until all three screens have been configured If you used the short cut to enter Puls...

Страница 21: ...s have been configured If you used the short cut to enter Manual Cycle Adjustment setup press and hold Mode button until Run Mode screen is displayed If you used the Menu button to navigate to Cycle A...

Страница 22: ...scroll over to next digit to the right If you pass your desired digit you can easily scroll back by continuously pressing RIGHT button Press ENTER to save the changes Continue this process until two...

Страница 23: ...Press RIGHT arrow to scroll over to next digit to the right If you pass your desired digit you can easily scroll back by continuously pressing RIGHT button Press ENTER to save the changes Continue thi...

Страница 24: ...ur desired digit you can easily scroll back by continuously pressing RIGHT button Press ENTER to save the changes Continue this process until all four screens have been configured If you used the shor...

Страница 25: ...t false triggering due to flickering high speed electrical switches a trigger delay time can be configured to delay the pump command The delay time unit of measure is seconds A two second delay time i...

Страница 26: ...side Anywhere on discharge side of MD 3 pump Wiring for sensor can be connected directly to a MD 3 pump Pump will stop pumping if sensor detects no flow A relay will then close allowing for remote al...

Страница 27: ...ty of less than 430 microsiemens will not be detected This system is capable of detecting presence of a large number of chemicals including Sodium Hypochlorite Chlorine Hydrochloric muriatic Acid Sodi...

Страница 28: ...sure input menu setting section 4 4 is switched to disabled while LOCAL ONLY is selected This menu setting will return to the previous setting when REMOTE ONLY or DISABLED is selected Input Setup Inpu...

Страница 29: ...00 Pump Run Stop Frequency Mode Cycle Timer Mode Dispensing Mode Manual Dosing Prop Dosing 0 10 VDC Mode Pulse Mode Disable FVS DFD Pump Run Stop 4 20mA Mode Manual Mode Frequency Mode Cycle Timer Mod...

Страница 30: ...ut mA Signal 0 Pump Output 4 mA 100 Pump Output 20 mA Signal Output Frequency Output Disable Done 4 20 mA Output Default 4 20 mA output settings shown Signal Output Disable Done 4 20 mA Output Default...

Страница 31: ...t assume responsibility for damage to pump that has been caused by chemical attack 6 2 Cleaning Pump Pump will require occasional cleaning especially Injection fitting Footvalve Strainer and pump head...

Страница 32: ...Page 32 MD 3 7 0 Exploded view and parts list 1 2 16 4 24 6 8 7 10 22 9 15 14 23 19 21 18 3 13 11 3 17 20 12 25...

Страница 33: ...TFE P 8 70001 346 CARTRIDGE VALVE EP 9 90002 272 PUMP HEAD 2 10 70004 115 PUMP HEAD FACE PLATE 2 11 90011 210 SCREW 10 32 X 2 0 SOC HD 16 12 90011 094 WASHER 10 4 13 90003 141 O RING 2 018 TFE P 4 900...

Страница 34: ...t 40 PSIg ML MIN 32 151 303 751 1508 2252 3041 RPM 1 3 10 19 48 95 142 190 GPH 0 5 2 4 4 8 11 9 23 9 35 7 48 2 Feed Rate at 90 PSIg ML MIN 25 145 278 694 1388 2107 2391 RPM 1 3 10 19 48 95 142 190 GPH...

Страница 35: ...Page 35 MD 3 Page intentionally left blank...

Страница 36: ...etering type of pump Should it become necessary to return the pump for repair or service you must attach information regarding the chemical used as some residue may be present within the unit which co...

Отзывы: