
To Input a Language:
1
Open the
App Drawer
2
Open
Settings
3
Open
System
4
Select
Locale
5
Select
Desired Language
!
English
!
Deutch
!
Español
!
Français
!
Portugues
6
Confirm by pressing
OK
12.5 Set Language
This setting is used to change the system language.
®
FLEXFLO A4
12.0 Settings
Page 44
Page 45
®
FLEXFLO A4
12.0 Settings
12.6 Pump Rotation Direction
This setting is used to change the rotational direction of pump. In most applications, the
tube will fail by developing a small leak in the outlet side (pressure side) of the tube
assembly. By reversing the roller rotation, the wear point in the tube is moved to the
opposite side to the pump tube assembly, increasing the life of the tube.
Important!
Changing the rotational direction of the pump reverses the inlet & outlet
sides.
IMPORTANT! Swap sides of the suction (inlet) and discharge (outlet) tubing/piping. There is no need to remove
the pump head cover.
NOTE: The pump tube will form a natural U-shaped curve. Do not attempt to install the pump tube against the
natural U-shape direction as damage to the tube can result.
To Change The Direction Of The Pump Rotation
1
Open the
App Drawer
2
Open
Settings
3
Open
System
4
Select
Pump Direction
5
Select
Desired Rotation
!
Clockwise
!
Counter Clockwise
6
Confirm by pressing
OK
Содержание FLEXFLO A4 Series
Страница 1: ...Peristaltic Metering Pump Series A4 ...
Страница 33: ...Page 33 FLEXFLO A4 10 0 Industrial Protocols 10 1 Control and Status Mapping for Profibus and Ethernet IP ...
Страница 55: ...Page 55 FLEXFLO A4 PAGE INTENTIONALLY LEFT BLANK ...
Страница 60: ... FLEXFLO A4 Page 60 PAGE INTENTIONALLY LEFT BLANK ...