11.1 Tube Info
This feature will display information regarding the tubing within the pump including:
!
Tube type
!
Tube installation date
!
Tube run time
!
Current maximum tube flow rate
To View The Tube Info:
1
Tap on the
Tube Info
text in the
top portion of the screen
2
Tube info will be displayed
3
Click “reset” to reset the tube
hours and revolutions
®
FLEXFLO A4
11.0 Tubing
Page 38
Page 39
®
FLEXFLO A4
11.0 Tubing
11.2 Tube Calibration
This feature allows the user to calibrate the pump’s indicated flow rate to the system
To Calibrate Your Tube:
1
On the home screen select the
Calibration Icon
2
Enter values:
!
Pump Speed (% Speed)
!
Run Time (seconds)
3
Select
Start
to begin
4
Select
Start
5
Enter the measured flow rate
into the field
6
Confirm by selecting
Save