background image

11

DC1 Setup 

The meter provides setup settings for each battery. To access these settings, first go to the Setup menu. Scroll to the desired battery label followed by 
Setup (such as DC1 Setup), then press 

Select

NOTE:

 Some settings may not be available for all batteries. 

The battery setup settings are described below. To make a change, scroll to setting and press 

Enter

. Press the 

LEFT 

 or  

RIGHT

   arrow buttons to 

view the available setting options. Press 

Enter

 to save the setting. Press 

Cancel

 to cancel a change. 

Enable 

To display the battery and its measurements, change this setting to 

ON

.  If enable is 

OFF

, the battery along with its measured values will not be displayed.

However, any associated alarm settings are still active. To de-activate the alarm, disable them in the Alarm Setup menu.

Set State-of-Charge (SoC) to FULL 

When this option is selected, the meter will consider the battery’s present State-of-Charge to be 

FULL

. To do this, scroll to 

Ok

 and press 

Select

.  

The screen will then return to the meter summary display. 

Battery Voltage  

This setting indicates the nominal voltage of the battery. The options are 12V, 24V, 36V, & 48V. This is automatically set by CZone.

Bat. Type  

This setting indicates the battery type. The available options are listed below: 
• FLStd - Standard Flooded Lead Acid Batteries 
• AGM - Standard Absorbed Glass Mat (AGM) Batteries 
• TPPL - Thin Plate Pure Lead AGM Batteries 
• GEL - Gel Batteries 
• FLRsv - Flooded Acid Reserve. Premium Flooded Lead Acid Batteries such as Rolls. 
• FLLoM - Flooded Low Maintenance. Sealed Flooded Lead Acid Batteries 
• FFFL - Firefly AGM Batteries

Bat. Capacity 

This setting indicates the capacity of the battery in amp-hours. The available range is 0A to 5000A. This is automatically set by CZone.

Full Chrg Volts

This setting indicates the voltage at which the battery is considered to be fully charged. The available range is 0.1 V to 70.0V. This should be set to the 
absorption voltage of your charger. If you don’t know the absorption voltage, you can figure the absorption voltage out by observing the output voltage 
of the charger once it enters the absorption phase. If you have multiple charging sources, set it to the higher absorption voltage. In extremely hot or cold 
temperatures the absorption voltage will need to be set slightly higher (for cold) or slightly lower (for hot). This is automatically set by CZone.

Example 1.

 In Blue Seas Systems’ P12 Charger this parameter is called the 

Absorb Voltage

. For lead acid values, the default value is 14.5 V DC. So in 

the M2 the 

Full Chrg Volts

 setting would be set slightly smaller than the 14.5 V (14.4 V).

Example 2.

 An unknown 10 A charger is connected to a partially discharged battery. Initially the meter should read approximately 10 A of charging current 

with the voltage increasing over time. At some point (around 14.5-14.8 V DC) the voltage should stabilize and the charging current will start decreasing. 
This voltage is the Full Charge Voltage. Set the 

Full Chrg Volts 

to 0.1 V below the observed voltage.

Содержание M2 Vessel Systems Monitor

Страница 1: ...Activation High and Low Voltage Current and Frequency Frequency Range 40Hz 90Hz Resolution 1Hz Power Range 0W 45kW Resolution 0W 9990W 10W Resolution 10kW 45kW 0 1kW Tanks Alarm High Low Levels Sender...

Страница 2: ...If an inverter is installed on the vessel its power leads must be disconnected at the battery before the meter is installed If an AC generator is installed on the vessel it must be stopped and render...

Страница 3: ...ount Bezel and Seal Mounting Ring and Nut Surface Mount Gasket Surface Mount Cover NOTE During installation use cover to align the bezel and gasket Flat Mount Surface Mount 360 Panel Mount PN 1525 Mou...

Страница 4: ...d 2 DC Pos connected These pins are used to provide power to the meter Connect pin 1 to ground and pin 2 to a 12V to 48V power source Pin 2 is also used to monitor the voltages on bank 2 The current d...

Страница 5: ...wisted to reduce the effects of interference from other electrical circuits AC Voltage The AC supply HOT Line or Ungrounded Conductor connection should be protected with a fast acting fuse of 0 25A to...

Страница 6: ...button beneath it The pop up menu will disappear after the first button is pressed 6 The menu system is a two dimensional matrix Pressing the UP or DOWN arrow buttons will transition the display betw...

Страница 7: ...ff exclusive of delays If no alarm is set this is the estimated time until the battery is completely empty Ah Remain Estimated remaining capacity of the battery Ah Used This is the actual Ah removed f...

Страница 8: ...to cancel any change NOTE A low voltage threshold cannot be set above a high voltage threshold Likewise a high voltage threshold cannot be set below the low voltage threshold The meter will automatica...

Страница 9: ...es the delay time When it reaches the delay timer an alarm will sound If the alarm condition goes away before the delay timer has been met the timer will count down until the timer is zero again For e...

Страница 10: ...either Imprl imperial or Mtrc metric These are the units used to specify the battery temperature and fluid volume Demo Mode With Demo Mode ON the meter displays factory programmed values for the Batt...

Страница 11: ...Batteries FLRsv Flooded Acid Reserve Premium Flooded Lead Acid Batteries such as Rolls FLLoM Flooded Low Maintenance Sealed Flooded Lead Acid Batteries FFFL Firefly AGM Batteries Bat Capacity This set...

Страница 12: ...or 10 hours We would then charge the battery and when we transition to float we might see 4 Ah Used So the charge efficiency would be 50 50 4 or 92 5 efficiency Temperature Specify the lowest temperat...

Страница 13: ...ngs first go to the Setup menu Scroll to the desired input The tank setup settings are described below To make a change scroll to setting and press Enter Press the LEFT or RIGHT buttons to view the av...

Страница 14: ...ank If the tank cannot be completely emptied the estimation works best if the tank is emptied with the sensor reading under 20 2 Set the tanks shape to Rect Rectangular 3 Set the tank to display perce...

Страница 15: ...te 5 The timer will start counting as the tank fills In addition the sender will indicate that the tank is getting filled When the tank reaches 100 the timer will automatically stop Press the Finish b...

Страница 16: ...nsport Protocol Connection Management 60928 ISO Address Claim 65240 ISO Commanded Address 126208 Request Group Function 126993 Heartbeat CZone Proprietary PGN s 65280 PGN_ZONE_COMMAND_MESSAGE Circuits...

Страница 17: ...USB micro cable into the back of the meter 6 While holding the 2nd button from the left connect the USB cable to a PC or a MAC b On a MAC 7 The computer will create a new drive on the computer Note I...

Страница 18: ...hould power up and you can navigate to Setup Version Info to verify the new SW Revision 4600 Ryzex Way Bellingham WA 98226 USA p 360 738 8230 f 360 734 4195 conductor bluesea com bluesea com 980033760...

Отзывы: