
BLEASE QUALITY DOCUMENT
BQD 084.V2
10/5/96
Page 2 of 3
Vent Measured
Ref
Description
BPM
TV(L)
Ratio
Del Freq
(BPM)
Del Ratio
Del TV/L
VTi
VTe
5.10
Vent Control
Functions
20C+20R
Adult Mode
Min Fresh Gas
6
6
10
20
20
20
40
60
90
1.0
1.0
1.0
0.6
0.4
0.2
0.1
0.2
0.1
1:1.0
1:Max
1:2.0
1:3.0
1:Min
1:2.0
1:2.0
1:1.0
1:2.0
-------------
-------------
-------------
-------------
-------------
1:
1:
1:
-------------
1:
1:
-------------
1:
1:
-------------
-------------
-------------
-----------
-----------
-----------
-----------
-----------
-----------
Ref
Operation\Description
Value 2
Value 1
Units
5.11
Expired Alarms, 1000 ml / 10 BPM, 1:2.0 Ventilator
Monitored MV (Value 2) & TV (Value 1)
MV/TV = 4L/200ml, Frq = 20 BPM
Decrease gas, MV low alarm operation <2.0L for 30 seconds
Increase gas, MV High alarm operation >6.0L for 30 seconds
-------------
-------------
P/F
P/F
VTi
VTe
-------------
-------------
L
-------------
P/F
P/F
5.12
Pat pr Alarms, 1000 ml / 12 BPM, 1:1.0
Delivered peak pressure (2)/vent monitored peak (1)
vent monitored mean
Pressure high (2) and low (1) limit alarm settings
-------------
-------------
-------------
-------------
cm H
2
O
cm H
2
O
cm H
2
O
5.13
Adult 50 cm H
2
O pressure limit, 1200 ml, 10 BPM, 1:2.0 -------------
cm H
2
O
5.14
Apnoea Alarms
Apnoea disconnection time (sens input), pressure limit = max
30
-------------
-------------
-------------
-------------
-------------
seconds
5.15
Vent Cycle Alarm , 20 BPM, disconnection time (sens input) -------------
seconds
5.16
Flow and waveform, TV 1L. 25 BPM, 1:2.0
Patient = C50, delivered TV
Paed bellows movement
50 ml bellows movement
200 ml bellows movement
-------------
-------------
-------------
-------------
-------------
-------------
-------------
-------------
L
-------------
ml
ml
5.17
Paed 20 cm H
2
O pressure limit, /200 ml, 30 BPM, 1:2.0
Bellows pause position
Sensor disconnection
-------------
-------------
-------------
cm H
2
O
ml
cm H
2
O
5.18
ASB Test
Trigger point 5, BPM 8500 (2), Timeter (1)
Trigger point 15 Breathing Stop
Alarm sound
Trigger point 5 Normal Running
Alarm silenced
-------------
-------------
-------------
-------------
-------------
-------------
BPM
P/F
P/F
P/F
P/F
Checked:-_______