Safety Information
39
Medical Equipment
If any other electrical equipment/accessory is connected to the system, the system
including
this equipment and/or accessory will become a medical system. Medical
systems must comply with EN/IEC 60601–1
,
ANSI/AAMI ES60601-1 or CAN/CSA
C22.2 No. 60601-1.
Non-Medical Equipment
WARNING
NME-w1
Follow the guidelines in EN/IEC 60601–1.
If you connect non-medical equipment (instruments that do not comply with safety
requirements for medical equipment, such as a video monitor, video recorder, endoscopic
camera control unit or other documentation device), this equipment must be placed
outside the patient environment (1.5m from the bed, for example). The equipment must
fulfill the relevant EN standard or other applicable national or international standard.
One of the following conditions must be fulfilled:
•
The system and other equipment are plugged into an external common isolation
transformer to control the leakage current during a ground connection fault.
or
•
The system is grounded with an additional safety ground connection (see “Additional
Protective Ground and Potential Equalization” on page 33).
If in doubt, contact your BK service representative.