44
Local Control
V = 300.00 V I = 1.000 A OFF
0.00 V 0.000 A
Analog Control
V = 300.0 V I = 1.00 A OFF
0.0 V 0.00 A
EXTERN
ENABLE
Local/Analog pin
state
Output
control
Display
NOT ACTIVE (by
Default)
NOT ACTIVE
Local
NO
DISPLAY
ACTIVE
High (5V)
Local
Low (0V)
Analog
Local/Analog State
Pin 21 can be used to indicate the current output control mode of the power
supply (Local or Analog state). In Local state, this pin will output a High
level (5V) signal, and in Analog State, this pin will output a Low level (0V)
signal.
Содержание XLN15010
Страница 1: ...Model XLN15010 XLN30052 XLN60026 High Power Programmable DC Power Supply USER MANUAL...
Страница 20: ...17 voltage is within the specified range...
Страница 111: ...108...
Страница 113: ...110...
Страница 114: ...111 22820 Savi Ranch Parkway Yorba Linda CA 92887 www bkprecision com 2013 B K Precision Corp Printed in Taiwan v042413...