Current Calibration
Follow the step by step instructions below for current calibration.
1.
Connect two terminals on the output of the power supply to two
ends of the current shunt, then connect the DVM to the sensor of
the current shunt to measure DC voltage, as shown below.
DVM
DC Voltage measurment
Current Calibration
DS30052
+
-
Current Shunt
+
-
Sense +
Sense -
2.
Press
from the calibration menu to enter the current
Содержание XLN15010