Harmonic Measurement Function
35
Function
Description
U(V)
Voltage
I(mA)
Current
P(mW)
Power
Q(mvar)
Reactive Power
S(mVA)
Apparent Power
PF()
Power Factor
𝜙
UI(
∘
)
Phase difference of k-order harmonic voltage and harmonic current
𝜙
UU(
∘
)
Phase difference of harmonic voltage U(k) and fundamental U(1)
𝜙
II(
∘
)
Phase difference of harmonic current I(k) and fundamental I(1)
U(%r) or U(%f) Voltage harmonic distortion
I(%r) or I(%f) Current harmonic distortion
P(%r) or P(%f) Power harmonic distortion
Table 6.2
Measurement Parameters
Figure 6.4
Harmonic Measurement Setup Menu
6.3.1 Parameters
Use the arrow keys to select the different parameters, and if numeric, the rotary knob to set the value, or the soft-keys.
THD Formal
Distortion factor calculation formula.
% r
displaying harmonics in the form of percentage to the overall current (voltage, power) amplitude
of all harmonics.
% f
displaying harmonics in the form of percentage to the fundamental wave current (voltage, power).
PLL Source
Select the PLL (Phase Locked Loop) source: U/I/OFF. Used for determining the fundamental
wave cycle as the reference for analysis of harmonic orders.
Serial
Harmonic sequence: whole sequence/odd sequence/even sequences
Order Max
Set the harmonic analysis orders (1-50). You can specify the harmonic measurement range. These
specified analysis orders are used for calculating the value of the distortion factor.