2.13 Ratio Setup
This setting has the effect of scaling the measured value of the current and voltage. For example, a 116V signal becomes
232V when the “Voltage Ratio” is set to 2. The Ranges are not effected by this setting, so the 150V range is still shown
and used. Use the arrow keys to move between parameters and digits. Use the knob to change the value of the selected
digit. To commit the setting change, press the
button. See
Figure 2.17
Ratio Setup Menu
2.14 External Current Sensor
More information to be added in future versions of this documentation.
When using an external current sensor, the power meter has more current range options.
Crest Factor
Ranges
EXT1 Crest Factor 3 (CF=3) 2.5 V, 5 V, 10 V
EXT1 Crest Factor 6 (CF=6) 1.25 V, 2.5 V, 5 V
EXT2 Crest Factor 3 (CF=3) 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V
EXT2 Crest Factor 6 (CF=6) 25 mV, 50 mV, 100 mV, 250 mV, 0.5 V, 1 V
Table 2.1
External Sensor Ranges