53
Execution Errors
An execution error indicates that the device could not execute a syntactically correct command, either
since the data were out of the instrument's range, or due to a device condition.
-200
Execution error
An attempt was made to RECALL the contents of an uninitialized
stored setting buffer.
-211
Trigger ignored.
The *TRG common command was ignored due to the device not
being in the correct state to execute the trigger.
-220
Parameter error.
A parameter is in the correct range, but conflicts with other
parameters.
-221
Settings conflict.
The parameter is out of range due to the current instrument
state.
-222
Data out of range.
-223
Too much data.
The arbitrary waveform memory limit has been exceeded.
-224
Illegal parameter value.
The parameter value must be selected from a finite list of
possibilities.
Device-Specific Errors
An error specific to the device occurred.
-315
Configuration memory lost.
Device memory has been lost.
-330
Self-test failed.
-350
Queue overflow.
Error codes have been lost due to more than 10 errors being
reported without being read.