BK Precision 2530B Скачать руководство пользователя страница 6

Trigger Adjustments 

 

1.

 

From the calibrator, setup a Sine signal with frequency of 200 Hz and amplitude of 3 Vpp.  Make 
sure there is 0 offset. 

2.

 

Input this signal into CH1 input and press AUTO button for automatic rescaling. 

3.

 

Press CH1 button and from the soft menu, select “Coupling” option to “AC”. 

4.

 

Press UTILITY button, then select “Do Self Cal” from Page 2/4. 

5.

 

With the black screen for self cal on display, do the following 

in order:

 

1.

 

Press CURSORS button once. 

2.

 

Press MEASURE button two times. 

3.

 

Press ACQUIRE button three times. 

6.

 

If you did it correctly, you will enter the calibration menu (shown below).  If you did not, redo the 
above steps again. 

 

7.

 

Select “Trig Adj” from the soft menu to enter the trigger adjustment menu.  There should be four 
soft menu options:  Source, Center, Delta, and Save. 

8.

 

Adjust the vertical and horizontal scale of the sine wave so that it fits about one and a half cycle of 
the waveform (see below) 

 

9.

 

We will first adjust the trigger center.  The trigger center is the point at which the signal is 
triggered at the horizontal trigger marker, aligned with the trigger level as illustrated below: 

Содержание 2530B

Страница 1: ...Model 2530B 2532B Calibration and Adjustment Procedure V111011...

Страница 2: ...Revision History Date Edition Revise Reason Revise Content Revised by 11 04 10 V1 0 None None Jason 11 10 11 V111011 Complete re edit All Jeremy...

Страница 3: ...complete users will need to lock access to the calibration menu Instructions to do this will be shown at the end of this document After unlocking access to the calibration menu the user must make note...

Страница 4: ...ill call this mean For positive DC 7 Next change the DC signal on the calibrator to 30 V 8 Make note of the measured Mean value again We will call this mean For negative DC 9 Now calculate and see if...

Страница 5: ...0 mV 150 0 mV 300 mV 20 0 mV 0 06VDC 0 06VDC 60 00 mV 60 00 mV 120 mV 10 0 mV 0 03VDC 0 03VDC 30 00 mV 30 00 mV 60 mV 5 00 mV 15mVDC 15mVDC 15 00 mV 15 00 mV 30 mV 2 00 mV 6mVDC 6mVDC 6 00 mV 6 00 mV...

Страница 6: ...ce 2 Press MEASURE button two times 3 Press ACQUIRE button three times 6 If you did it correctly you will enter the calibration menu shown below If you did not redo the above steps again 7 Select Trig...

Страница 7: ...14 Now select center from the soft menu and use the universal knob to increase or decrease the value until you catch the critical point between where the signal is triggered and not triggered You wil...

Страница 8: ...oints signal is no longer triggered and will move across the screen o 3 12V 7 04V 2 5 08 V This will be the current trigger center level o Turn the LEVEL knob until trigger level is set to 5 08 V leve...

Страница 9: ...and observe the trigger range If the trigger range exceeds the waveform display range meaning that the signal is still triggered when trigger level is beyond the waveform display level then the delta...

Страница 10: ...ument needs to do a self calibration prior to use To do this press UTILITY and go to Page 2 4 and select Do Self Cal With nothing connected at all the inputs This is very important as it will affect c...

Страница 11: ...TILITY button 2 Select Do Self Cal on Page 2 4 of the menu and you will see the black screen that prompts for initiating self calibration 3 At this time do the following in order 1 Press DEFAULT SETUP...