Calibration Procedure 

 
To calibrate the 2530B or 2532B Digital storage oscilloscope, user must first unlock access to calibration 
menu. 
 

How to Unlock Access to Calibration Menu 

 

1.

 

Unplug all attached inputs, power off the instrument, and restart. 

2.

 

Press UTILITY button. 

3.

 

Then, go to Page 2/4 and select “Do Self Cal” from the menu options. (You should have a black 
screen with the phrase “Disconnect Everything from All Inputs” at the top) 

4.

 

Now do the following 

in order:

 

1.

 

Press the CURSORS button once. 

2.

 

Press the DEFAULT SETUP button once. 

3.

 

Press the SET TO 50% button once. 

4.

 

Press the 5

th

 menu selection button once (This is the button directly above the PRINT 

button on the right side of the display) 

5.

 

Press the AUTO button once. 

6.

 

Press the FORCE button once. 

5.

 

If you follow the above steps correctly, you should be returned back to the main display with the 
UTILITY menu opened.  If not, repeat steps 2-4 again. 

6.

 

The instrument is now unlocked for access to the calibration menu.  After re-adjustments are 
complete, users will need to lock access to the calibration menu.  Instructions to do this will be 
shown at the end of this document. 

 
After unlocking access to the calibration menu, the user must make note of the following 

FIRST 

before 

continuing below to make the appropriate adjustments. 
 

NOTE: 

-

 

The instrument MUST run (to warm up) for AT LEAST 15 minutes BEFORE any 
adjustments are made. 

-

 

Power to the instrument must be connected at all times during the course of adjustments. 

 

Voltage Adjustments 

 

1.

 

From your calibrator instrument, setup a 30 V DC signal and connect it to CH1 input of the 
oscilloscope, then press AUTO button once.  (This will automatically adjust and align the signal 
on the display) 

2.

 

Press the UTILITY button, and go to Page 2/4 to select “Do Self Cal” from the menu options.  
(You should have a black screen with the phrase “Disconnect Everything from All Inputs” at the 
top) 

3.

 

Now, do the following 

in order:

 

1.

 

Press CURSORS button once. 

2.

 

Press MEASURE button two times. 

3.

 

Press ACQUIRE button three times. 

4.

 

If you did it correctly, you will enter the calibration menu (shown below).  If you did not, redo the 
above steps again.  Select “Volt Adj” from the soft menu to enter the voltage adjustment menu. 

Содержание 2530B

Страница 1: ...Model 2530B 2532B Calibration and Adjustment Procedure V111011...

Страница 2: ...Revision History Date Edition Revise Reason Revise Content Revised by 11 04 10 V1 0 None None Jason 11 10 11 V111011 Complete re edit All Jeremy...

Страница 3: ...complete users will need to lock access to the calibration menu Instructions to do this will be shown at the end of this document After unlocking access to the calibration menu the user must make note...

Страница 4: ...ill call this mean For positive DC 7 Next change the DC signal on the calibrator to 30 V 8 Make note of the measured Mean value again We will call this mean For negative DC 9 Now calculate and see if...

Страница 5: ...0 mV 150 0 mV 300 mV 20 0 mV 0 06VDC 0 06VDC 60 00 mV 60 00 mV 120 mV 10 0 mV 0 03VDC 0 03VDC 30 00 mV 30 00 mV 60 mV 5 00 mV 15mVDC 15mVDC 15 00 mV 15 00 mV 30 mV 2 00 mV 6mVDC 6mVDC 6 00 mV 6 00 mV...

Страница 6: ...ce 2 Press MEASURE button two times 3 Press ACQUIRE button three times 6 If you did it correctly you will enter the calibration menu shown below If you did not redo the above steps again 7 Select Trig...

Страница 7: ...14 Now select center from the soft menu and use the universal knob to increase or decrease the value until you catch the critical point between where the signal is triggered and not triggered You wil...

Страница 8: ...oints signal is no longer triggered and will move across the screen o 3 12V 7 04V 2 5 08 V This will be the current trigger center level o Turn the LEVEL knob until trigger level is set to 5 08 V leve...

Страница 9: ...and observe the trigger range If the trigger range exceeds the waveform display range meaning that the signal is still triggered when trigger level is beyond the waveform display level then the delta...

Страница 10: ...ument needs to do a self calibration prior to use To do this press UTILITY and go to Page 2 4 and select Do Self Cal With nothing connected at all the inputs This is very important as it will affect c...

Страница 11: ...TILITY button 2 Select Do Self Cal on Page 2 4 of the menu and you will see the black screen that prompts for initiating self calibration 3 At this time do the following in order 1 Press DEFAULT SETUP...

Отзывы: