How to Lock Access to Calibration Menu
Once all adjustments are made and self calibration is completed, the instrument must be set back to lock
access to the calibration menu. This will bring the instrument back to normal operation mode. Follow the
below instructions to do this.
1.
Press UTILITY button.
2.
Select “Do Self Cal” on Page 2/4 of the menu and you will see the black screen that prompts for
initiating self calibration.
3.
At this time do the following
in order:
1.
Press DEFAULT SETUP button once.
2.
Press SAVE/RECALL button two times.
3.
Press the UTILITY button three times.
4.
If this is done correctly, you should not be able to access the calibration menu by selecting
UTILITY button, select “Do Self Cal”, press CURSORS button once, MEASURE button twice,
ACQUIRE three times.