22
MAINTENANCE
WARNING
The following instructions are for use by qualified personnel only. To
avoid electrical shock, do not perform any servicing other than
contained in the operating instructions unless you are qualified to do
so.
FUSE REPLACEMENT
If the fuse blows, the CV or CC indicators will not light and the
power supply will not operate. The fuse should not normally open
unless a problem has developed in the unit. Try to determine and
correct the cause of the blown fuse, then replace only with a fuse of
the correct rating as listed below. The fuse is located on the rear panel
(see Fig. 2).
OPERATION
FUSE VALUE
TYPE
120 V
3.5 A
SLOW BLOW
220/230/240 V
1.75 A
SLOW BLOW
Table 1. Fuse Values.
LINE VOLTAGE CONVERSION, INTERNATIONAL
UNITS
The primary winding of the power transformer is tapped to
permit operation from 120/220/230 or 240 VAC, 50/60 Hz
line voltage. Conversion from one line voltage to another is
done by a simple position change of the fuse receptacle
underneath the line cord.
ADJUSTMENTS
This unit was accurately adjusted at the factory before
shipment. Readjustment is recommended only if repairs have
been made in a circuit affecting adjustment accuracy, or if
you have a reason to believe the unit is out of adjustment.
However, adjustments should be attempted only if a
multimeter with an accuracy of ±0.5% DCV or better is
available (
B+K Precision
Model 2703B or equivalent).
Содержание 1743B
Страница 11: ...OPERATING INSTRUCTIONS 11 Figure 3 A and B Grounding Possibilities...
Страница 12: ...12 OPERATING INSTRUCTIONS Figure 3 C and D Grounding Possibilities...
Страница 17: ...OPERATING INSTRUCTIONS Figure 8 Connecting Two Power Supplies in Series 17...
Страница 18: ...18 OPERATING INSTRUCTIONS Figure 9 Connecting Two Power Supplies in Parallel...
Страница 24: ...MAINTENANCE Figure 11 Location of Adjustments 24...