About the device
Operating instructions eS10
10
38034911005 en
2
2.4.1
Position of identification plates
The main identification plate in accordance with OIML R76-1 is located on the right side of
the housing. The metrological information is located on an additional plate which is near
the window close to the display.
1
Main identification plate in accordance
with OIML R76-1
2
Identification plate with max, min and e
Fig. 5: Position of identification plates
2.5
Intended use
The device is intended to be used as a compact scale / terminal in an industrial environ-
ment.
–
The device may only be used in industrial or commercial fields.
–
The device may not be used in potentially explosive areas.
–
The device must only be transported using appropriate means of transportation and if
it was secured accordingly.
–
No unauthorized constructional modifications must be applied to the device.
–
The device may only be used for verifiable transactions if the software is operating
correctly and is in a secure state. The software IDs must be correctly displayed and
checked by the operator daily. The internal safety mark must not be damaged.
2.6
Protective mark locations
2.6.1
Markings and labels
All markings and labels that are required according to the valid EC directives are located
on the identification plate. The symbol for conformity evaluation (initial verification) is af-
fixed to the measuring device.
2.6.2
Safety plaques
The scale adjustment data are protected by a safety clip under the lead seal. The connec-
tor for the load receptor is also located under the sealing cover. The fixing screw of the
lead seal is secured against removal by a yellow seal mark.
For load receptors with analog load cells, the cable connection cabinet is also provided
with a seal mark.