Right LED red: RIP belt(s) error. Check RIP belt(s) as well as the connection
between Respiro and the patch (press studs).
Note 1. It is possible to have errors in several sensors simultaneously.
Note 2. Error states are indicated when the equipment is worn on body.
Note 3. If some of the LEDs are green, the sensor in question is ok.
Table 6 Respiro 4 sensor configuration LED indications in error states
4 sensor configuration (2 RIP belts with ECG), 2 LEDs on for 1 second cycling from left to right
Left LED yellow, center green: Pulse oximeter is connected but finger is not
inserted in the sensor. Check connection.
Left LED green, center LED red: Nasal cannula error. Check connection.
Left LED red, center LED green: Pulse oximeter error. Check connections.
Left and center LED red: Pulse oximeter and nasal cannula error. Check con-
nections.
Center LED green, right LED red: RIP belts error. Check RIP belts.
Center LED red, right LED green: ECG error. Check electrode.
Center and right LED red: ECG and RIP belts error. Check RIP belts and elec-
trode.
Note 1. Error states are indicated when the equipment is worn on body and recording is ongo-
ing.
Note 2. If some of the LEDs are green, the sensor in question is ok.