Respiro
16 (48)
Operating Instructions
© Bittium Biosignals 2022
All three LEDs turn successively green: Battery charge 90-94 %.
All LEDs static green: Battery charge 95-100 %.
2.5.4
Power button functions
Respiro device’s power button has the following functions:
•
A press of approx. 3 seconds: Respiro power on.
•
After power on a press of approx. 8 seconds after which Respiro vibrates
once: Recording start.
•
A press of approx. 3 seconds during recording: Recording end. Respiro vi-
brates 3 times.
•
Short press (<3 seconds): Patient event marker set indication during record-
ing.
•
A press of approx. 12 seconds: Respiro device power off.
2.6
Before recording
Make sure that the Respiro
device’s battery charge status is sufficient (at least
two
green LEDs are blinking successively according to Table 2. Note! If scheduled
recording is used all three indicator LEDs must be blinking successively to ensure
sufficient battery charge for the recording) for the planned recording and that the
pulse oximeter’s batteries are good and have a sufficient charge.
Change pulse ox-
imeter’s
batteries if required, that is when there are two segments left in the bat-
ter
y symbol on pulse oximeter’s display
.