5
Chapter 3
Installation
This chapter provides information for on-site requirements, unpacking,
inspection, and preparing the load for use.
Unpacking and Inspection
1. Carefully inspect shipping container for signs of damage.
2. Do one of the following:
z
If the shipping container is damaged, do not unpack the unit. Immedi-
ately notify the shipping carrier and Bird Electronic Corporation.
z
If the shipping container is not damaged, unpack the unit. Save ship-
ping materials for repackaging.
3. Inspect unit for visual signs of damage.
Note:
If there is damage, immediately notify the shipping carrier
and Bird Electronic Corporation.
Setting Up
z
Before first using the load, get a resistance baseline for future mainte-
nance. Refer to "Inspecting the DC Resistance" on page 16 for details.
z
Remove the shipping plug from the load and replace it with the vent
plug. Refer to Figure 3 for pictures of the plugs.
Mounting
Place the load in a dry, dust and vibration free environment. Do not use out-
doors or in areas of condensing humidity. Allow at least 12" (30 cm) of clear-
ance on all sides of the load.
Bird 889x Loads are equipped for either portable use or fixed installation. The
mounting brackets on the front and rear faces have four mounting slots
arranged in a 4
1
⁄
2
" x 20
23
⁄
32
" rectangle (114.3 x 526.3 mm). Use a screw with a
3
⁄
8
" (9.53 mm) diameter max.
WARNING
The vent plug must be used at all times when the unit is operating or
cooling. Failure to do so could result in an explosion or severe burns.
CAUTION
This load is designed for operation in a horizontal position only, with the
vent plug up. Do not use in any other manner.
深圳市新朗普电子科技有限公司
www.xLp17.com Email:[email protected]
TEL:0755-83986300 FAX:0755-88850515
QQ:55178055 285480356
深圳市新朗普电子科技有限公司
www.xLp17.com Email:[email protected]