4.6.4.1 Placement for Groin Pain
in One Location
Use the U-set:
Place the round
Pain Site Electrode on the groin
directly over the painful area. The
comfortable location for the large
rectangular Dispersive Electrode is
horizontally across the lumbar region
on the lower back (out of view in the
photo to the right).
Body Position:
For groin treatments,
lying supine is the most desirable
(comfortable) position for the patient
during the treatment. Legs should
be kept straight.
4.6.4.2 Placement for Groin Pain in Two Locations
Use the B-set:
Place each round Pain Site Electrode on the groin directly over
each painful area. If pubic hair is in the location of the painful area, then the
hair must be shaved so that the electrode can make good electrical contact
with the skin. Leave at least 1.0 inch of spacing between the two Primary Round
Electrodes. There is no limit on the maximum distance the electrodes can be
placed apart from on another. If the electrodes are in close proximity to the
patient’s genitals, use a barrier like a towel between the electrodes and genitals.
Body Position:
For groin treatments, lying supine is the most desirable
(comfortable) position for the patient during the treatment. Legs should be
kept straight.