Test Procedure
1. Prepare the
75% Test Solution
by mixing 15 mL of the 100% Test Solution with 5
mL of the PBS Solution.
2. Prepare the
50% Test Solution
by mixing 10 mL of the 100% Test Solution with 10
mL of the PBS Solution.
3. Carefully pipette the three solutions into a new 96-well microplate:
l
150 μL of the 100% Test Solution into all wells of columns 1 and 2
l
150 μL of the 75% Test Solution into all wells of columns 3 and 4
l
150 μL of the 50% Test Solution into all wells of column 5 and 6
4. Create a Gen5 experiment based on the
Synergy H1 Abs Test 3
protocol and read
the plate.
l
Save the experiment. Refer to the instructions on page
to perform calculations
and determine pass/fail.
l
Troubleshooting tips are provided on page
.
l
Test descriptions are provided on page
.
118 | Chapter 8: Instrument Qualification Procedures
BioTek Instruments, Inc.
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...