Specifications
Page 39 of 42
Epoch 2
Instructions for Use
1771011
Revision
A
BioTek Instruments, Inc.
Specifications
General Specifications
Microplates
The
Epoch 2
accommodates standard 6-, 12-, 24-, 48-, 96-, and 384-well microplates with 128 x 86
mm geometry and the
Take3
Micro-Volume Plate. If using
Gen5
, the
Take3
Trio
Micro-Volume plate
and
BioCell
are also supported.
Maximum plate height: 20.32
mm
Hardware and Environmental
Weight:
Fully configured weight ~11.3
kg
Dimensions:
Touchscreen models: 31.8
cm
x 32.4
cm
x 39.4
cm
Non-touchscreen models: 21.6
cm
x 32.4
cm
x 39.4
cm
Operational Temperature
Range:
Touchscreen models: 18
°C
to 30
°C
Non-touchscreen models: 18
°C
to 40
°C
Humidity:
10% to 85% relative humidity (non-condensing)
Power Consumption:
Powered from an external 120
W
(minimum), 24
VDC
power supply
compatible with 100-240 volts
AC
@50-60
Hz
.
Temperature Control
Range:
Maximum incubation temperature 65
°C
. Minimum setpoint must be
ambient room temperature plus 4
°C
. For incubation setpoints > 60
°C
,
ambient room temperature must be at least 22
°C
.
Uniformity:
+/- 0.5
°C
at 37
°C
, tested with
Innovative Instruments, Inc
. temperature
test plate.