background image

20

TurboVap

®

 96 Dual User Manual | © Biotage 2021

Maintenance

Clean the Manifolds

Warning

 

»

Ensure not to bend any of the nozzles. If a nozzle is bent, it is 

recommended that the manifold is replaced.

 

»

The nozzles may have come into contact with hazardous 

samples. Avoid contact. 

Clean the manifolds regularly. The frequency of cleaning 
should be determined by the needs of the laboratory, but it is 
recommended that the manifolds should be cleaned when:

 

»

a manifold may have been contaminated,

 

»

a manifold has been in contact with 
harsh solvents or additives,

 

»

a nozzle is obstructed,

 

»

before long term storage.

The most thorough way to clean the needle assemblies is to 
remove the manifolds and clean them outside of the instrument. 
However, in-situ rinsing can be performed as a quick 
alternative.

The nozzles are made of 316 stainless steel (EN 1.4404), and the 
head assembly is made of anodized aluminum (aluminum, EN 
AW-6082). Always consider the material of the instrument parts 
when selecting cleaning agent.

Clean a Manifold in Place

This method requires a suitable container that can be filled with 
cleaning agent and that fits on the plate heater. A well collection 
plate (2 mL or larger) is recommended, see "Consumables" on 
page 24

Note: 

Ensure that the well collection plate fits the manifold, i.e. 

for a 24-nozzle manifold use a 24-well collection plate.

1.  Fill the container with an appropriate cleaning solvent. The 

solvent is dictated by analyte solubility and the following is 
recommended:
a.  For polar analytes chose solvent combinations with a 

high aqueous content, e.g., an 80/20 solution of  water 
and methanol.

b.  For lipophilic analytes chose solvent combinations with 

a high organic solvent content, e.g., an 80/20 solution of 
methanol and water.

2.  Place the container on the plate heater under the manifold.
3.  Close the door and lift the container so that the nozzles are 

submerged in the desired cleaning solvent. (Do not start an 
evaporation run.) 

4.  Keep the nozzles submerged in the solvent for  5-10 minutes.

Note:

 

For polar analytes it is recommended to repeat step 1 to 4 

with a high organic solvent combination after the high aq rinse.

5.  Lower the container and remove it. 
6.  Ensure that the nozzles are dry before initiating an 

evaporation run. Either wait until they are dry, or perform an 
evaporation run with an empty cabinet and unheated gas. 
(25 °C/77 °F) 

Clean a Manifold Outside of the Instrument

This method requires the manifold to be removed from the 
instrument and placed in a sonicator. Additionally, the method 
requires a suitable container that can be filled with solvent. 
A well collection plate (2 mL or larger) is recommended, see 
"Consumables" on page 24 .

1.  Remove the manifold as described in "Replace the Manifold" 

on page 11

.

2.  Prepare a container as in "Clean a Manifold in Place" on 

page 20.

3.  Place the container in a sonicator. Ensure that the water level 

is below the edge of the container. Water in the container 
will lead to contamination.

4.  Carefully lower the manifold into the container and sonicate 

for 5-10 minutes.
a.  If contamination is present, the times and solvent 

variants may need to be adjusted. 

Содержание TurboVap 96 Dual

Страница 1: ...Patent Pending Biotage 2021 Title Truncated if Neccessary Page 1 TurboVap 96 Dual User Manual...

Страница 2: ...8 Define System Settings 8 Common Settings 9 Evaporation Settings 10 Network and Firmware 10 System Information 10 Warnings in the Software 11 Prepare the System 11 Replace the Manifold 11 Set the Pr...

Страница 3: ...s manifolds with 24 48 or 96 nozzles to allow for processing of up to 96 samples in each cabinet The left and right cabinet can perform simultaneous evaporation runs with varying size well plates vary...

Страница 4: ...he door is closed the evaporation run will resume To determine the correct plate height for an evaporation run see Determine Optimal Evaporation Conditions on page 6 Evaporation To provide consistency...

Страница 5: ...the Manifold on page 11 according to the chosen well plate Available manifolds TurboVap manifold with 24 gas nozzles TurboVap manifold with 48 gas nozzles TurboVap manifold with 96 gas nozzles Differe...

Страница 6: ...radation of analytes may be observed Do not allow samples or solvents to boil The evaporation time of samples decreases with increasing sample cleanliness The presence of sample detritus or particulat...

Страница 7: ...plate Usually setting the temperature of the plate heater 0 20 C above the gas temperature allows for enhanced evaporation uniformity Use of the plate heater minimizes cross contamination of samples w...

Страница 8: ...ifying and calibrating the touch screen 1 Mode Mode selection Press to select mode 2 Common Settings Press to change common settings 3 Evaporation Settings Press to change evaporation settings 4 Netwo...

Страница 9: ...as temperature Tick the box to wait for the gas temperature to reach the set value before starting an evaporation Wait for plate temperature Tick the box to wait for the plate temperature to reach the...

Страница 10: ...ress the Reset Settings button to reset all settings in Common settings Evaporation Settings and Network and Firmware to their default values Calibrate Touch Press the Calibrate Touch button to calibr...

Страница 11: ...ure 17 Ensure that the manifold is held in position while the last screw is removed 3 With a steady grip remove the manifold Note Do not hold the manifold by the nozzles as this might bend the nozzles...

Страница 12: ...Time Mode Evaporate for a set period of time 4 Method Mode Evaporate using a predefined method 5 Split to left and right OR Join left and right The left and right cabinets operate independently OR The...

Страница 13: ...splay shows that the left side is ready to start the orange Start button is active and the right side is running an evaporation the blue Pause button and the red Stop button are active To perform a Ti...

Страница 14: ...ode In the Join left and right mode the left and right cabinets operate the same evaporation run with the same parameters see Figure 23 Figure 23 Temp Step Method 96 run in Join left and right mode Du...

Страница 15: ...utton to enter the desired name see Evaporation Parameters on page 15 6 When done press Save Note Once a method has been saved its previous settings cannot be recovered 7 In the Methods view press OK...

Страница 16: ...ers view where the evaporation parameters or the method name can be edited or a method can be copied Figure 29 Press and hold the Plate height button then move the lift by pressing the Up and Down arr...

Страница 17: ...an alarm sounds If operating each cabinet separately then the LED lights in each cabinet will turn green independently The Evaporation Finished dialog appears on the screen see Figure 34 The alarm co...

Страница 18: ...l risk for burn damage Open the cabinet door The well plate will descend Carefully remove the well plate or rack Be careful not to come in contact with the nozzles See Figure 35 Note The lift will ret...

Страница 19: ...ly dampened with isopropanol or ethanol 3 When the system has been cleaned connect the power cord and turn on the system Clean the Interior of the System Warning Never clean the surface where the mani...

Страница 20: ...following is recommended a For polar analytes chose solvent combinations with a high aqueous content e g an 80 20 solution of water and methanol b For lipophilic analytes chose solvent combinations wi...

Страница 21: ...plied by Biotage Incorrect fuses will void the warranty of the system and creates a potential fire hazard and a risk of personal injury 1 Turn off the system and disconnect the power cord 2 Using a fl...

Страница 22: ...into Storage Warning Ensure not to bend any of the nozzles If a nozzle is bent it is recommended that the manifold is replaced 1 Remove and clean the manifolds as described on page 20 2 Place the man...

Страница 23: ...er Temperature on page 7 Check that the sample is not left too long on the plate heater after the evaporation is completed Check that the gas flow is not too high and causing splashing see Determine t...

Страница 24: ...crew 1 Consumables Part No Description Qty 121 5201 Collection Plate 350 uL Square 50 121 5202 Collection Plate 1 mL Square 50 121 5203 Collection Plate 2 mL Square 50 121 5213 Collection Plate 2 mL R...

Страница 25: ...25 Notes Notes...

Страница 26: ...art Number 418001 B 2021 Biotage All rights reserved No material may be reproduced or published without the written permission of Biotage Information in this document is subject to change without noti...

Отзывы: