411829-L, Maintenance
April 2012
Page 5-9
5
6. Ensure that the drain tube for the solvent tray is not damaged and is safely connected to the
Drain port at the rear of the system. The other end shall be inserted into a separate waste
reservoir.
Figure 5-6. Drainage of Spillage on the Solvent Tray
WARNING
To avoid being struck by the collection arm, keep your hands out of range
of the collection arm while the homing routine runs in step 7 below.
7. Once you have located and taken care of the leakage, reconnect the system and turn it on. The
collection arm moves through its homing routine and the system boots to the system’s main
menu.
8. Check all tubes and connections for leaks using the prime function; see “Prime the System” on
page 4-33. Prime with water or another suitable solvent.
9. If the problem persists, press
End Prime
, shut down the system (see page 4-1), disconnect
the power cord from the power outlet, and contact Biotage 1-Point Support (see contact details
on page 7-1).
5.10.4 Assembling Tubes Correctly
WARNING
Shut down the system before replacing any tubing.
Use only tubing, nuts, and ferrules supplied by Biotage.
All external tubing on the Isolera systems except for the tubing inside the collection arm can be
replaced by the user; a list of spare parts is available at www.biotage.com.
If using an Isolera system with flangeless tubing, please notice that proper sealing is only achieved
with the ferrule oriented as shown in Figure 5-7, with the tapered portion of the ferrule facing
toward
the nut. The ferrule should be placed near the end of the tube as shown below.
Figure 5-7. Flangeless Tubing with Properly Oriented Ferrule
If using super flangeless tubing, ensure the ferrules are fitted onto the tube ends as described in
the instructions supplied with the ferrules.
When using (super) flangeless tubing, use caution when finger tightening fittings to prevent
stripped threads or crushed ferrules.
Drain port
Drain tube
Содержание Isolera Four
Страница 1: ...Isolera User Manual...
Страница 66: ...411829 L Operation April 2012 Page 4 16 4 Figure 4 10 Tube Connections...
Страница 114: ......
Страница 115: ......