411829-L, System Settings
April 2012
Page 3-5
3
3.3.5 Enable the
λ
-All Detection Mode and TLC to Step Gradient Editor
To enable the
λ
-all detection mode and the TLC to Step Gradient editor, you need a Spektra license.
To purchase a Spektra license, please contact your local representative. Note that the license is not
available for Isolera Prime systems.
To install a Spektra license for the
λ
-all detection mode and the TLC to Step Gradient editor:
1. Save the Spektra license file in a “biotage/isolera/” directory on a USB memory device.
2. Connect
the memory device to the USB port located underneath the touch screen.
3. Select the
General
tab in the right-hand panel.
4. Press
Install License
in the
Licenses
field.
5. If the license is valid,
License OK
appears in the
Advanced Features
text box.
6. Remove the USB memory device.
7. Restart the system:
a. Press
Main Menu
and then
Shut Down
.
b. When the message
It is now safe to turn off the system
appears on the screen, turn off
the system. The power switch is located underneath the touch screen.
c. Turn on the system.
3.4 Configure a Network Connection
With your system connected to the network, it is possible to:
• Print purification reports to a network printer with postscript support; see “Set Up a Printer and
Auto Print of Reports” on page 3-6.
• Receive a report by e-mail when a purification has been completed; see “Set Up Auto Send of
• Receive an e-mail when user interaction is required during a purification, e.g. when the system
run out of empty collection vessels.
• Make your reports and methods available to other users by saving them in a share folder on
your network; see “Set Up File Sharing and Auto Save of Reports” on page 3-7.
• Back up your database and save the system configuration and logs in a share folder on your
network; see “Back Up and Restore the System’s Database” and “Export the System
Configuration and Logs” on page 3-12.
With the system connected to your network, it is also possible to perform the following tasks
through a standard web browser (through the Isolera Remote Viewer feature):
• Create a new method. If desired, the method can be based on a previous purification found
in the Results view.
• Check the status of the system and the progress of a purification.
• Instruct the system to start and stop collecting fractions by clicking the
Collect All
button
( = fractions are collected).*
• Instruct the system to switch to a new collection vessel by clicking the
New Fraction
button.*
• Instruct the system to start and end an isocratic segment by clicking the
Hold
button
( = isocratic hold is enabled).*
• Instruct the system to pause the purification in progress by clicking the
Pause
button. For
safety reasons, it is only possible to pause and not resume a run from your office.*
• Instruct the system to set the current UV level to 0 (zero) AU by clicking the
UV Zero
button.*
• View, print, and export purification results.
* You must be logged in as the user defined in the purification run.