411829-L, System Settings
April 2012
Page 3-11
3
4. To enable or disable air flush, select the
Air Flush
text box and select
To Waste
or
Off
. This
feature is not available with Isolera Prime.
5. To change the solvent or solvent mix to be used for the line, detector, and/or air flush, select
the corresponding text box and select the desired solvent or solvent mix:
•
Method
– The system will flush with the weakest or strongest solvent used in the run.
•
System
– The system will flush with the weakest or strongest solvent mounted on the system.
•
Gradient
– The system will flush with the solvent mix used in the start or the end of the
gradient. (Not available for air flush.)
•
Channel x
– The system will flush with the solvent mounted on the specified channel (1, 2,
3, or 4); see “Assign Solvents to the Solvent Inlets” on page 4-32. Note that Isolera One,
Four, and LS have four solvent inlets while Isolera Prime has two.
3.6.8 Configure Collection and Fractionation
Refer to the “Collection and Fractionation” appendix on page A-1 for details.
1. Select the
Runtime
tab in the right-hand panel.
2. Select the
Start Slope Enable
text box and enter the value when start slope is enabled.
3. Select the
Shoulder/Valley Slope Disable
text box and enter the level when shoulder and
valley fractionation is disabled.
4. Select the
Valley Slope
text box and enter the level when valley fractionation occurs.
3.7 Maintenance
3.7.1 Export the System Configuration and Logs
If requested by 1-Point Support™ to send in your system’s log files, follow the instructions below.
The system configuration and logs can be saved on a USB memory device connected to the USB
port at the front of the system or in a share folder on your network (if file share has been set up;
see page 3-7).
1. Select the
Maintenance
tab in the right-hand panel.
2. If you want to save the system configuration and logs on a USB memory device:
a. Connect
the memory device to the USB port located underneath the touch screen.
b. Press
Export
in the
System Configuration and Logs
field.
3. If you want to save the system configuration and logs in the specified share folder, press
Export
in the
System Configuration and Logs
field.
4. The system configuration and logs are saved as a zip file at \biotage\isolera\logs\. Send the zip
file to 1-Point Support; see “Contact Information” on page 7-1.
3.7.2 Restore the System Configuration
The system configuration contains all the settings available in the System mode except for the
internal detector calibration (which is stored on the detector), the local printer (if connected), and
the date and time.
NOTE
Only restore the system configuration when instructed to do so by Biotage.
Do not restore the system configuration on your Isolera system using a
system configuration for another Isolera system.
If you have problem with the restore, please contact 1-Point Support.