411829-L, System Settings
April 2012
Page 3-12
3
To restore the system configuration e.g. after an update of the system's operating system:
1. Select the
Maintenance
tab in the right-hand panel.
2. Connect
the USB memory device containing the zip file with the desired system configuration to
the USB port located underneath the touch screen. Note that the zip file must be saved at
\biotage\isolera\logs\ on the USB memory device.
3. Press
Restore
Configuration
in the
System Configuration and Logs
field.
4. In the
Select Zip File
dialog, select the zip file and press
OK
. The
Restore from USB Device
dialog opens.
5. To confirm restore, press
Restore
.
6. When the
Restore Successful
dialog appears, press
OK
. The
Restart Required
dialog opens.
7. Read the instructions in the dialog on how to restart the system and then press
OK
.
8. Restart the system.
9. If you have a USB printer connected to the system, set up the printer as described in “Set Up a
Printer and Auto Print of Reports” on page 3-6.
10. Verify that the calibration of the touch screen and the fraction collector is working properly. If
the system needs to be recalibrated; see “Calibrate the Touch Screen” on page 3-12 and
“Calibrate the Fraction Collector” on page 3-13.
3.7.3 Back Up and Restore the System’s Database
To back up the database:
1. Select the
Maintenance
tab in the right-hand panel.
2. If you want to save the backup on a USB memory device:
a. Connect a memory device to the USB port located underneath the touch screen.
b. Press
Back
Up
in the
Back Up and Restore Database
field.
3. If you want to save the backup in the specified share folder, press
Back Up
in the
Back Up
and Restore Database
field. The backup file is saved at \biotage\isolera\backup\.
To restore the database:
1. Select the
Maintenance
tab in the right-hand panel.
2. If the backup file you want to use to restore the database is saved on a USB memory device
(at \biotage\isolera\backup\):
a. Connect the memory device to the USB port located underneath the touch screen.
b. Press
Restore
in the
Back Up and Restore Database
field.
3. If the backup is available in the specified share folder (at \biotage\isolera\backup\), press
Restore
in the
Back Up and Restore Database
field.
4. In the
Select Backup File
dialog, select the backup file and press
OK
. The
Restore from USB
Device/Share Folder
dialog opens.
5. To confirm restore, press
Restore
.
3.7.4 Calibrate the Touch Screen
If the touch screen needs to be recalibrated, use the following procedure.
1. Select the
Maintenance
tab in the right-hand panel.
2. Press
Calibrate
in the
Touch Screen
field. Read and follow the instructions that appear on the screen.
Содержание Isolera Four