Biotage Extrahera Classic Скачать руководство пользователя страница 8

6

Biotage

®

 Extrahera

 Installation and Safety | © Biotage 2020

Technical Specifications

Note:

 The system contains corrosion sensitive parts.  

Best practice is to avoid sustained continuous exposure 
to acidic and basic vapors by always removing the solvent 
reservoirs when the system is not in use, and cleaning the 
system following usage. Usage of strong acids (e.g. TFA and 
TCA) and inorganic acids (e.g. nitric, sulfuric, hydrochloric acid, 
and perchloric acids) in the solvent pumps is not supported 
at any concentration.

System Requirements

Ambient 

temperature

Operating temperature: 18°C to 32°C

Storage and transportation temperature:  

-25°C to 60°C 

Humidity

10% to 90% RH 

Electrical supply

100 to 240 VAC, 50/60 Hz  

Fuses

T4A at the power inlet (2 required) 

Maximum power 

consumed 

300 VA

Gas supply

The gas applied to the samples (i.e. the gas 

connected to the 

N

2

 port) must be according 

to ISO 8573-1, classification 1-6-1. The filter 

unit that is supplied with the system delivers 

according to ISO 8573-1, class 1-x-1. If the 

gas applied is not according to class x-6-x, 

an air dryer needs to be ordered separately 

(P/N 416441SP).

AIR

 port: 5 L/min (0.18 cubic feet/min), 

 

±

 0.2 bar (0.6 

±

 0.02 MPa; 87 

±

 3 psi) for  

a fully populated extraction plate/rack and  

lower to approximately 4 bar (0.4 MPa; 58 psi) 

when 50% of the plate/rack is populated and  

3 bar (0.3 MPa; 44 psi) when 25% of the 

plate/rack is populated.

N

2

 port: 1 L/min (0.04 cubic feet/min) 

during processing and 70 L/min (2.5 cubic 

feet/min) during plate dry, 6 ± 0.2 bar 

(0.6 ± 0.02 MPa, 87 ± 3 psi) 

Vacuum supply

Partial pressure range: < 500 mbar (50 kPa; 

7 psi)

Flow rate: ≥ 8.3 L/min (0.29 cubic feet/min)

 

Weight

75 kg (165 lbs) 

Dimensions 

(WxDxH)

Without touch screen: 610 x 510 x 730 mm 

(24.0" x 22.1" x 28.7") 

With touch screen: 860 x 570 x 730 mm  

(33.8" x 22.4" x 28.7") 

Max sound level

65 dB(A) 

Liquid Handling 

Solvent inlets

5

Solvents are pumped into five 25 mL reservoirs 

inside the system using the five solvent pumps.

It is possible to have an extra solvent rack with 

five 25 mL or 100 mL reservoirs. These have to 

be filled manually. 

The dead volume is 4 mL for both the 25 and 

100 mL reservoir, and the 100 mL reservoir has 

a maximum fill volume of 90 mL.

 

Waste outlet

Pipetting pump

50 to 1000 µL 

At 50 µL: ±2.0% accuracy and 1.0% CV

 

At 100 µL: ±2.0% accuracy and 1.0% CV

 

At 500 µL: ±1.5% accuracy and 1.0% CV

 

At 1000 µ: ±1.0% accuracy and 1.0% CV 

Processing  

format

96-well extraction plates, 96-array plates for 

1 and 2 mL wells, 96-position extraction racks 

for 1 mL (tabless) columns (A format), 48-well 

extraction plates, 48-position extraction racks 

for 3 mL (tabless) columns (B format), and 

24-position extraction racks for 1, 3, and 6 mL 

(tabless) columns (A, B, and C format) 

Processing 

pressure 

0 to 5 bar

Interfaces

Touch screen

12.1" 

Ethernet LAN

Complies with IEEE 802.3 (ANSI 8802-3) 

USB

USB 2.0

Technical Specifications

Содержание Extrahera Classic

Страница 1: ...Biotage Extrahera Installation and Safety...

Страница 2: ...l Specifications 6 System Requirements 6 Liquid Handling 6 Interfaces 7 Clog Detection 8 Safety 8 Intended Use 8 Education Training and Competence 8 Warranty and Liability 8 Service 8 Safety Features...

Страница 3: ...unded outlet Vacuum supply Partial pressure range 500 mbar 50 kPa 7 psi Flow rate 8 3 L min 0 29 cubic feet min Connection The vacuum tube supplied with the system has a 6 mm 0 24 outer diameter and 4...

Страница 4: ...in position 5 on the working area c Press Maintenance in the main menu d Press Flush Solvent Inlets e Enter the flush volume for S1 25 mL is required to empty the solvent inlet line and pump of liquid...

Страница 5: ...e height of the feet There is a built in spirit level in the bottom of the system in the front left corner See Figure 3 Figure 3 The built in spirit level in the bottom of the system in the front left...

Страница 6: ...system and your network 27 If you have disconnected the touch screen reconnect it by reversing the instructions in step 14 Note Do not mix up the Ethernet cables Connect according to the labeling on t...

Страница 7: ...and the S4 solvent feeder tube left port 5 S5 The S5 solvent inlet line right port and the S5 solvent feeder tube left port Figure 13 6 7 8 Waste connections located on the back wall inside the system...

Страница 8: ...s populated and 3 bar 0 3 MPa 44 psi when 25 of the plate rack is populated N2 port 1 L min 0 04 cubic feet min during processing and 70 L min 2 5 cubic feet min during plate dry 6 0 2 bar 0 6 0 02 MP...

Страница 9: ...d Heptane Good Hexane Poor Isooctane Good Isopropanol Excellent Methanol Excellent Pentane Poor Tetrahydrofuran THF Poor Toluene Excellent Water Excellent Methyl tert butyl ether MTBE Poor IPA 30 Exce...

Страница 10: ...h hazardous biological radioactive or toxic samples and or solvents prior to any service being performed When returning equipment to Biotage this should be done in accordance with the material return...

Страница 11: ...e suitable lifting equipment when moving the package The system weighs 75 kg 165 lbs The four lifting handles supplied with the system must be used when moving the system Four persons are required whe...

Страница 12: ...ulations All samples and waste should be treated as potentially biohazardous Always follow local and national safety regulations related to storage handling and disposal of chemicals biological sample...

Страница 13: ...ic Equipment WEEE Products falling under the scope of the WEEE Directive are identified with a crossed over wheelie bin symbol on the product label as indicated to the left To forward a product for re...

Страница 14: ...the back of this document or visit our website www biotage com Manufacturer Biotage GB Limited United Kingdom for Biotage Sweden AB Contact Us Biotage Sweden AB Box 8 SE 751 03 Uppsala SWEDEN Visiting...

Страница 15: ......

Страница 16: ...Part Number 414157 J 2020 Biotage All rights reserved No material may be reproduced or published without the written permission of Biotage Information in this document is subject to change without no...

Отзывы: