6
Biotage
®
Extrahera
™
Installation and Safety | © Biotage 2020
Technical Specifications
Note:
The system contains corrosion sensitive parts.
Best practice is to avoid sustained continuous exposure
to acidic and basic vapors by always removing the solvent
reservoirs when the system is not in use, and cleaning the
system following usage. Usage of strong acids (e.g. TFA and
TCA) and inorganic acids (e.g. nitric, sulfuric, hydrochloric acid,
and perchloric acids) in the solvent pumps is not supported
at any concentration.
System Requirements
Ambient
temperature
Operating temperature: 18°C to 32°C
Storage and transportation temperature:
-25°C to 60°C
Humidity
10% to 90% RH
Electrical supply
100 to 240 VAC, 50/60 Hz
Fuses
T4A at the power inlet (2 required)
Maximum power
consumed
300 VA
Gas supply
The gas applied to the samples (i.e. the gas
connected to the
N
2
port) must be according
to ISO 8573-1, classification 1-6-1. The filter
unit that is supplied with the system delivers
according to ISO 8573-1, class 1-x-1. If the
gas applied is not according to class x-6-x,
an air dryer needs to be ordered separately
(P/N 416441SP).
AIR
port: 5 L/min (0.18 cubic feet/min),
6
±
0.2 bar (0.6
±
0.02 MPa; 87
±
3 psi) for
a fully populated extraction plate/rack and
lower to approximately 4 bar (0.4 MPa; 58 psi)
when 50% of the plate/rack is populated and
3 bar (0.3 MPa; 44 psi) when 25% of the
plate/rack is populated.
N
2
port: 1 L/min (0.04 cubic feet/min)
during processing and 70 L/min (2.5 cubic
feet/min) during plate dry, 6 ± 0.2 bar
(0.6 ± 0.02 MPa, 87 ± 3 psi)
Vacuum supply
Partial pressure range: < 500 mbar (50 kPa;
7 psi)
Flow rate: ≥ 8.3 L/min (0.29 cubic feet/min)
Weight
75 kg (165 lbs)
Dimensions
(WxDxH)
Without touch screen: 610 x 510 x 730 mm
(24.0" x 22.1" x 28.7")
With touch screen: 860 x 570 x 730 mm
(33.8" x 22.4" x 28.7")
Max sound level
65 dB(A)
Liquid Handling
Solvent inlets
5
Solvents are pumped into five 25 mL reservoirs
inside the system using the five solvent pumps.
It is possible to have an extra solvent rack with
five 25 mL or 100 mL reservoirs. These have to
be filled manually.
The dead volume is 4 mL for both the 25 and
100 mL reservoir, and the 100 mL reservoir has
a maximum fill volume of 90 mL.
Waste outlet
1
Pipetting pump
50 to 1000 µL
At 50 µL: ±2.0% accuracy and 1.0% CV
At 100 µL: ±2.0% accuracy and 1.0% CV
At 500 µL: ±1.5% accuracy and 1.0% CV
At 1000 µ: ±1.0% accuracy and 1.0% CV
Processing
format
96-well extraction plates, 96-array plates for
1 and 2 mL wells, 96-position extraction racks
for 1 mL (tabless) columns (A format), 48-well
extraction plates, 48-position extraction racks
for 3 mL (tabless) columns (B format), and
24-position extraction racks for 1, 3, and 6 mL
(tabless) columns (A, B, and C format)
Processing
pressure
0 to 5 bar
Interfaces
Touch screen
12.1"
Ethernet LAN
Complies with IEEE 802.3 (ANSI 8802-3)
USB
USB 2.0
Technical Specifications
Содержание Extrahera Classic