background image

2

Biotage

®

 Extrahera

 Installation and Safety | © Biotage 2020

Installation

Positive 

pressure unit

The gas applied to the samples (i.e. the gas 

connected to the 

N

2

 port) must be according to 

ISO 8573-1, classification 1-6-1. This is essential to 

prevent sample contamination and general fouling 

of the pressure unit. The filter unit that is supplied 

with the system delivers according to ISO 8573-1, 

class 1-x-1. If the gas applied is not according 

to class x-6-x, an air dryer needs to be ordered 

separately (P/N 416441SP). 

AIR

 port requirements: 5 L/min (0.18 cubic feet/

min), 6 

±

 0.2 bar (0.6 

±

 0.02 MPa; 87 

±

 3 psi) for a 

fully populated extraction plate/rack and lower to 

approximately 4 bar (0.4 MPa; 58 psi) when 50%  

of the plate/rack is populated and 3 bar (0.3 MPa; 

44 psi) when 25% of the plate/rack is populated.

N

2

 port requirements: 1 L/min (0.04 cubic 

feet/min) during processing and 70 L/min 

(2.5 cubic feet/min) during plate dry, 6 ± 0.2 bar 

(0.6 ± 0.02 MPa, 87 ± 3 psi)

Connection: The gas tube supplied with the 

system has a 6 mm (0.24") outer diameter and 

4 mm (0.16") inner diameter. 

External fire 

protection

External fire protection should be installed 

according to local regulations for equipment 

operating unattended.

Move the System

Warning

 

»

Before moving the system, please read and observe the safety 

requirements in "Safety Requirements" on page 9.

If you need to move the system within the laboratory or between 
laboratories in the same building, follow the instructions below. 
If you need to ship the system, please contact Biotage

®

 1-Point 

Support

 for instructions.

1.  Prepare the new site according to the site requirements 

on page 1

2.  Clean the waste tubing as described in the user manual 

(P/N 414158). 

3.  Empty the system of liquids by flushing it with air: 

a.  Remove the solvent inlet lines from their bottles and 

place them in an empty, clean bottle.

b.  Ensure that you have five empty solvent reservoirs in  

the solvent rack in position 

5

 on the working area. 

c.  Press 

Maintenance

 in the main menu.

d.  Press 

Flush Solvent Inlets…

e.  Enter the flush volume for S1. 25 mL is required to  

empty the solvent inlet line and pump of liquid. 

f.  Press 

Flush

 for S1. 

g.  When you have finished flushing the solvent inlet line 

with air, repeat steps e to f for the other solvent inlet 
lines (S2-S5). 

4.  Remove the solvent racks in position 

and

 6

 (if used) on the 

working area

 

and empty the reservoirs.

5.  Remove all accessories and consumables inside the system, 

i.e. the pipette tip waste bin, plates, racks, etc.

6.  Turn off the vacuum pump or, if your system is connected to 

another vacuum source, close the valve. 

7.  Shut down the system by pressing 

Shut Down

 in the main 

menu or login view (GLP) and then 

Yes

 to confirm.

8.  When the message saying that it is safe to turn off the system 

appears on the screen, turn off the system. The mains switch 
is located on the right hand side at the rear of the system.

9.  Unplug the power cord from the power outlet.
10. Turn off the gas supply/supplies, and disconnect the 

pressure regulators and the gas tubes from 1) the supply/
supplies and 2) the 

AIR

 and 

N

2

 ports located on the right 

hand side at the rear of the system. 

11.  Disconnect the waste outlet tube from the waste reservoir 

and empty the waste reservoir.

12. Clean the interior of the system and the accessories as 

described in the user manual (P/N 414158).

13. Clean the touch screen and the exterior of the system, 

using a soft and clean cloth. The cloth can be dry or lightly 
dampened with a neutral detergent or alcohol.

14. If you need to remove the touch screen (computer):

a.  Adjust the ball mount angle so that the rear of the touch 

screen can be accessed; see Figure 1

b.  Remove the lower panel by unscrewing the six screws 

using a Philips #2 screwdriver.

c.  Disconnect the power connector by unscrewing its 

locking ring. 

d.  Disconnect the two Ethernet cables and any USB 

devices.

e.  Remove the touch screen by loosening the ball mount.

Note:

 The touch screen must be placed on a flat and level 

surface to avoid damage. 

Figure 1.

 The rear of the touch screen (computer).

Содержание Extrahera Classic

Страница 1: ...Biotage Extrahera Installation and Safety...

Страница 2: ...l Specifications 6 System Requirements 6 Liquid Handling 6 Interfaces 7 Clog Detection 8 Safety 8 Intended Use 8 Education Training and Competence 8 Warranty and Liability 8 Service 8 Safety Features...

Страница 3: ...unded outlet Vacuum supply Partial pressure range 500 mbar 50 kPa 7 psi Flow rate 8 3 L min 0 29 cubic feet min Connection The vacuum tube supplied with the system has a 6 mm 0 24 outer diameter and 4...

Страница 4: ...in position 5 on the working area c Press Maintenance in the main menu d Press Flush Solvent Inlets e Enter the flush volume for S1 25 mL is required to empty the solvent inlet line and pump of liquid...

Страница 5: ...e height of the feet There is a built in spirit level in the bottom of the system in the front left corner See Figure 3 Figure 3 The built in spirit level in the bottom of the system in the front left...

Страница 6: ...system and your network 27 If you have disconnected the touch screen reconnect it by reversing the instructions in step 14 Note Do not mix up the Ethernet cables Connect according to the labeling on t...

Страница 7: ...and the S4 solvent feeder tube left port 5 S5 The S5 solvent inlet line right port and the S5 solvent feeder tube left port Figure 13 6 7 8 Waste connections located on the back wall inside the system...

Страница 8: ...s populated and 3 bar 0 3 MPa 44 psi when 25 of the plate rack is populated N2 port 1 L min 0 04 cubic feet min during processing and 70 L min 2 5 cubic feet min during plate dry 6 0 2 bar 0 6 0 02 MP...

Страница 9: ...d Heptane Good Hexane Poor Isooctane Good Isopropanol Excellent Methanol Excellent Pentane Poor Tetrahydrofuran THF Poor Toluene Excellent Water Excellent Methyl tert butyl ether MTBE Poor IPA 30 Exce...

Страница 10: ...h hazardous biological radioactive or toxic samples and or solvents prior to any service being performed When returning equipment to Biotage this should be done in accordance with the material return...

Страница 11: ...e suitable lifting equipment when moving the package The system weighs 75 kg 165 lbs The four lifting handles supplied with the system must be used when moving the system Four persons are required whe...

Страница 12: ...ulations All samples and waste should be treated as potentially biohazardous Always follow local and national safety regulations related to storage handling and disposal of chemicals biological sample...

Страница 13: ...ic Equipment WEEE Products falling under the scope of the WEEE Directive are identified with a crossed over wheelie bin symbol on the product label as indicated to the left To forward a product for re...

Страница 14: ...the back of this document or visit our website www biotage com Manufacturer Biotage GB Limited United Kingdom for Biotage Sweden AB Contact Us Biotage Sweden AB Box 8 SE 751 03 Uppsala SWEDEN Visiting...

Страница 15: ......

Страница 16: ...Part Number 414157 J 2020 Biotage All rights reserved No material may be reproduced or published without the written permission of Biotage Information in this document is subject to change without no...

Отзывы: