15. Attach the four lifting handles to the system; see Figure 2

The handles are supplied with the system.

Figure 2.

 The four lifting handles supplied with the system must be used 

when moving the system.

16. The system weighs 75 kg (165 lbs). Carefully lift the system 

using the four lifting handles and place it on a trolley that 
can support the weight of the system. Four persons are 
required when lifting the system. 

17.  Move the trolley and the rest of the equipment (the pipette 

tip waste bin, waste reservoir, vacuum pump (if used), 
solvent bottles, etc.) to the new location.

18. Carefully lift the system and place it in a well-ventilated 

fume hood or on a bench. Four persons are required when 
lifting the system. Place the system so that the mains switch 
is easy to access. 

19. If the system is not placed in a fume hood, connect the outlet 

of the Extrahera top ventilation to a ventilation system: 
a.  Connect an air duct adapter from Biotage to the outlet of 

the Extrahera top ventilation. 

b.  Connect a ventilation tube between the air duct adapter 

and the exhaust. The outer diameter of the air duct 
adapter is 159.0 ± 0.3 mm (6.26" ± 0.01") and the inner 
is 152.7 ± 0.3 mm (6.01" ± 0.01"), i.e. the 160 mm tube 
should be mounted outside the air duct adapter and the 
6" tube on the inside of the air duct adapter. Ensure that 
the tube cannot be blocked. 

20. It is important that the system is not tilting backward.  

Level the system by adjusting the height of the feet. There is 
a built-in spirit level in the bottom of the system, in the front 
left corner. See Figure 3.

Figure 3.

 The built-in spirit level in the bottom of the system, in the front 

left corner.

21. Place the solvent bottles on the side of the system and 

insert the solvent inlet lines into the correct bottles.  
Ensure to use appropriate caps to prevent harmful solvent 
vapors from escaping and the contents from being spilled. 

22. Place the waste reservoir on the side of the system and 

connect the waste outlet tube and vacuum to it; see Figure 4
If using a vacuum pump, ensure that the vacuum pump 
fumes are directed into a proper ventilation system. 

Figure 4

Connect to the upper port of the 

waste valve, which is located on 

the left side of your system

Connect to 

the vacuum 

source

Approx. 7 cm

.

 The setup of the waste kit.

23. Before connecting air and/or nitrogen to the system, 

the pressure head must be removed to protect it from 
contamination:
a.  Remove the two screws holding the pressure head to the 

pressure unit; see Figure 5

Figure 5.

 The two screws attaching the pressure head to the pressure unit. 

b.  Remove the pressure head by pulling it down and then 

pulling it out.

Note:

 If you have a system with only one gas tube 

connected to the pressure head, you have to tilt the 
pressure head horizontally to be able to pull it out.

c.  Disconnect the gas tubes (or tube) from the pressure 

head by pushing in the collar of each connector against 
the fitting and pulling the tubing out; see Figure 6.

d.  Put the pressure head on a clean and lint-free surface. 

Figure 6.

 System with two gas tubes connected to the pressure head.

Содержание Extrahera Classic

Страница 1: ...Biotage Extrahera Installation and Safety...

Страница 2: ...l Specifications 6 System Requirements 6 Liquid Handling 6 Interfaces 7 Clog Detection 8 Safety 8 Intended Use 8 Education Training and Competence 8 Warranty and Liability 8 Service 8 Safety Features...

Страница 3: ...unded outlet Vacuum supply Partial pressure range 500 mbar 50 kPa 7 psi Flow rate 8 3 L min 0 29 cubic feet min Connection The vacuum tube supplied with the system has a 6 mm 0 24 outer diameter and 4...

Страница 4: ...in position 5 on the working area c Press Maintenance in the main menu d Press Flush Solvent Inlets e Enter the flush volume for S1 25 mL is required to empty the solvent inlet line and pump of liquid...

Страница 5: ...e height of the feet There is a built in spirit level in the bottom of the system in the front left corner See Figure 3 Figure 3 The built in spirit level in the bottom of the system in the front left...

Страница 6: ...system and your network 27 If you have disconnected the touch screen reconnect it by reversing the instructions in step 14 Note Do not mix up the Ethernet cables Connect according to the labeling on t...

Страница 7: ...and the S4 solvent feeder tube left port 5 S5 The S5 solvent inlet line right port and the S5 solvent feeder tube left port Figure 13 6 7 8 Waste connections located on the back wall inside the system...

Страница 8: ...s populated and 3 bar 0 3 MPa 44 psi when 25 of the plate rack is populated N2 port 1 L min 0 04 cubic feet min during processing and 70 L min 2 5 cubic feet min during plate dry 6 0 2 bar 0 6 0 02 MP...

Страница 9: ...d Heptane Good Hexane Poor Isooctane Good Isopropanol Excellent Methanol Excellent Pentane Poor Tetrahydrofuran THF Poor Toluene Excellent Water Excellent Methyl tert butyl ether MTBE Poor IPA 30 Exce...

Страница 10: ...h hazardous biological radioactive or toxic samples and or solvents prior to any service being performed When returning equipment to Biotage this should be done in accordance with the material return...

Страница 11: ...e suitable lifting equipment when moving the package The system weighs 75 kg 165 lbs The four lifting handles supplied with the system must be used when moving the system Four persons are required whe...

Страница 12: ...ulations All samples and waste should be treated as potentially biohazardous Always follow local and national safety regulations related to storage handling and disposal of chemicals biological sample...

Страница 13: ...ic Equipment WEEE Products falling under the scope of the WEEE Directive are identified with a crossed over wheelie bin symbol on the product label as indicated to the left To forward a product for re...

Страница 14: ...the back of this document or visit our website www biotage com Manufacturer Biotage GB Limited United Kingdom for Biotage Sweden AB Contact Us Biotage Sweden AB Box 8 SE 751 03 Uppsala SWEDEN Visiting...

Страница 15: ......

Страница 16: ...Part Number 414157 J 2020 Biotage All rights reserved No material may be reproduced or published without the written permission of Biotage Information in this document is subject to change without no...

Отзывы: