Biotage ELSD-1080
(i.e. STANDBY or RUN) you wish the unit to start-up in from the
sub menu screen (see Section 3.3.4, page 33). Switch off the
detector and wait for at least 10 seconds before switching the
unit back on, in order for the change to take affect. If
RUN
mode is selected as the desired Power mode, then the detector
will use the operating parameters stored in memory. In the
unlikely event that the detector encounters a fault during
power-up the unit will automatically switch to
STANDBY
mode.
3.4.9
Data Output Rate (Hz)
The rate at which the ELSD-1080 outputs data can be selected
from the sub-menu screen. A 10Hz output rate is suitable for
flash applications (the default value is 10Hz).
3.5
Controlling the Biotage ELSD-1080 Detector
Using the Standalone Software Utilities
These Windows™-based graphical interfaces offer a second and
complementary level to total detector control. An intuitive
single control panel provides simplistic control as well as a
comprehensive monitoring system. Operational parameters can
be easily manipulated, saved or loaded by using the flexible
Methods Editor enabling rapid set-up and custom method
archiving.
A general description of the features available through the
software utilities is given below, but for a more comprehensive
guide on how to use the control software, please refer to the
on-line help supplied with the software.
3.5.1
Control Software-Overview
Control of the ELSD-1080 is achieved through the simple
control window. Within the control window, the status of the
detector is displayed, together with options to control the
various detector parameters. The control window is effectively
divided into three main sections:
1.
Control Window
f
View the current mode of operation
f
Change the current mode of operation between
STANDBY and RUN
f
Change the current Method of operation between a
stored custom method and the default method ‘XXX’
f
Change detector parameters directly via method ‘XXX’
2.
Status window
f
Displays the actual values against the set point
for:
i.
Evaporator Temperature
ii.
Nebulizer Temperature
iii.
Gas Flow
Publication date: July 2010
37
Содержание ELSD-1080
Страница 4: ...Biotage ELSD 1080 This page is intentionally left blank 4 Publication date July 2010 ...
Страница 8: ...Biotage ELSD 1080 This page is intentionally left blank 8 Publication date July 2010 ...
Страница 44: ...Biotage ELSD 1080 This page is intentionally left blank 44 Publication date July 2010 ...
Страница 54: ...Biotage ELSD 1080 This page is intentionally left blank 54 Publication date July 2010 ...
Страница 56: ...Biotage ELSD 1080 This page is intentionally left blank 56 Publication date July 2010 ...
Страница 58: ...Biotage ELSD 1080 This page is intentionally left blank 58 Publication date July 2010 ...