18. Error Codes
Error code
Explanation
Measurement Error
E1: aborted measurement
The measurement timeout is over, the measurement had
to be aborted (the patient was moving).
E2: manually interrupted
The measurement was stopped by pressing a button. The
display differs from others: “OFF” on the LCD.
E3: battery rundown
The batteries exhausted during measurement.
E4: batteries replaced
The batteries were replaced during measurement.
E8: pressure limit exceeded
The pressure in the pneumatic system exceeded the
preset pressure limit
E9: temporary disturbance
External electric signal - static discharge - disturbed the
operation of the device.
Cuff Error
E31: cuff missing or loose
There was no cuff connected to the device or the cuff was
too loose on the patient’s arm.
E32: cuff tubing clogged
The cuff is clogged or the rubber tube is broken.
E33: cuff leaking or loose
There is a hole in the cuff or it is rather loose on the
patient’s arm.
E34: cuff not on arm
The patient did not wear the cuff.
Device Error
E90: device error
The device could not measure due to a hardware error.
E99: device error
The device does not start automatic measurement until
the next programming.
19. Product Warranty Information
MONITOR WARRANTY.
The main monitor unit will be free from defects in materials and workmanship under normal
use and service for a period of two (2) years from the date of receipt. This warranty covers the monitor only. This
warranty does not cover any accessories that might come with the monitor.
ACCESSORIES WARRANTY.
The non-disposable accessories delivered with the monitor unit will be free from
defects in materials and workmanship under normal use and service for a period of one (1) year from the date of
receipt. This warranty does not cover disposable accessories, packaging materials, accumulators and batteries, cuffs,
or any of their components.
CUFF WARRANTY.
The cuff - will be free from defects in materials and workmanship under normal use and service