Never place the monitor in a disinfecting or sterilizing machine!

13. Maintenance

Verification of the pressure measurement accuracy is recommended biannually.
All the devices are covered by a two-year warranty

 under general warranty conditions of BIOS Medical, 

see relevant topic. This warranty does not cover any malfunction or defects arising from improper use, the use 

of inadequate accessories, accident, theft, or use of the device outside operating environmental specifications or 

intended measurement range. Removing the closing label from the back side of the device voids this warranty.

There are no user serviceable parts inside the recorder;

 they contain high complexity electronic and fine 

mechanical components. If you have any problems, please call the BIOS Blood Pressure Hotline . All consequences of 

improper servicing are the sole responsibility of the user. Contact BIOS Medical for more information.

14. Disposal

Each Ambulatory Blood Pressure Monitor includes an internal NiCd coin cell which falls under the category of 

hazardous waste and should be disposed with proper care. The other parts of the device should be handled as normal 

electronic waste at roll-out. Used batteries may also fall under the category of hazardous waste and should be 

disposed with proper care.

15. Indication & Contraindications

Indications

-  Suspected white-coat hypertension

-  Suspected nocturnal hypertension

-  To establish dipper status

-  Resistant hypertension

-  Elderly patient

-  As a guide to antihypertensive drug treatment

-  Type 1 diabetes

-  Hypertension of pregnancy, including pre-eclamptic patients

-  Evaluation of hypotension

-  Autonomic failure

Contraindications

-  Non-cooperative patients, unconscious or otherwise incapable patients

-  Patients requiring urgency/emergency cardiac care

-  Patients with coagulation disturbances

-  Patients with serious mobility or other impairments without supervision

-  Children without supervision, or children younger than 8 years

-  Though the blood pressure measurement algorithm used in the monitors has been found to function properly on 

patients with atrial fibrillation or other common arrhythmias, the oscillometric blood pressure measurement method 

is generally recommended for use only with special caution in patients with arrhythmias, Parkinson’s disease or 

other diseases involving tremors.

16. Accessories List

- Ambulatory blood pressure monitor

- USB optical with USB driver cable

Содержание ABP-01

Страница 1: ...09 24 18 For 24 or 48 hour continuous monitoring Ambulatory Blood Pressure Monitor Instruction Manual ABP 01 ...

Страница 2: ... Monitor 4 Fit the Patient with the Monitor 5 Retrieve the Data 6 Customize Review Print Data 7 Manual Programming 8 Patient Information 9 Cuffs 9 1 Application 9 2 Apply the cuff 9 3 Connect the hose to the monitor 10 Batteries 11 Safety Concern 12 Cleaning and Protection 13 Maintenance 14 Disposal 15 Indications Contraindications 16 Accessories List 17 Technical Specifications 18 Error Codes 19 ...

Страница 3: ...and control The ambulatory blood pressure monitor incorporates an algorithm validated to BHS British Hypertension Society and AAMI Association for the Advancement of Medical Instrumentation protocols About the Ambulatory Blood Pressure Monitor START button LCD display EVENT button DAY NIGHT button Interface Connection Battery Compartment Sticker With Serial Number Cuff Connection ...

Страница 4: ...asurement is cancelled by pressing a button Diastolic value of just completed measurement mmHg Heart symbol blinking measurement in progress mmHg Night mode time is displayed moon sign is lit LCD check all segments are displayed Battery voltage display 2 37V The crossed battery symbol warns of low battery Communication with a personal computer Blood pressure measurement initiated kPa Pumping for m...

Страница 5: ...nds to check the LCD Press and hold the start button to light up all segments of the LCD to check if they all work correctly 4 To check battery voltage press and hold for more than 5 seconds but less than 10 seconds The voltage for fully charged batteries should be over 2 5V 2_50 on the LCD 1 2B Event Button Functions 1 To set a patient event marker press shortly Typical cause for this use is taki...

Страница 6: ...ting the USB optical cable to the PC The USB driver is installed together with the software in the absence of which the PC will not recognize the interface Locate a USB port on your computer and connect the USB driver Then connect the USB optical cable into the USB driver IMPORTANT It is very important to always plug the USB driver into the computer first and then plug the USB optical cable into t...

Страница 7: ...Adjust the active passive morning and special if applicable period and interval times b Press OK to confirm 5 Click on the RELEASE button the software will speak to the monitor and set up all the parameters from the software to the ambulatory monitor 6 Exit out of the software and unplug the monitor from the computer 4 Fit the Patient with the Monitor 1 Apply a proper size cuff to the patient s no...

Страница 8: ...the three plans undecided cuff size 300 mmHg pressure limit LCD display enabled manual day night shift disabled Daytime starts at 6 00 while nighttime starts at 22 00 special session is disabled Patient data can be selected or created later in the database How To Program Manually Press and hold the START and the DAY NIGHT buttons simultaneously Measurement frequency of the measurement plans will b...

Страница 9: ...he cuff should be removed from the arm immediately and disconnected from the monitor Such occurrence should be reported to the physician later after the monitoring session e Press any button to stop a blood pressure measurement Should the blood pressure measurements cause bloodshots torpidity or pain in the hand the cuff should be removed from the arm immediately and disconnected from the monitor ...

Страница 10: ... tube pointing downwards the advantage is that the patient can wear a loose jacket over the cuff When properly applied the end of the sleeve the one closer to the tube should fall in the indicated range 9 3 Connect the hose to the monitor Connect the air connector of the cuff into the air connector socket of the device by turning it clockwise with slight pressure NOTE Take care to avoid blocking t...

Страница 11: ...atteries as they may not cover the power needs of the monitor and they may damage the monitor or they may contain acidic electrolytes which may leak and corrode electronic components Never use batteries damaged in any way b Do not start a new monitoring session with low batteries It is strongly recommended to use freshly charged or new batteries with each patient so that batteries do not run down ...

Страница 12: ... decisions of a qualified physician they do not automatically provide a diagnosis of any kind During the evaluation of recorded blood pressure values possible artefacts due to external disturbances motion artefacts and electrical noise should be observed and handled with caution See chapter Cuffs for more information 12 Cleaning Protection The ambulatory blood pressure monitor is not specially pro...

Страница 13: ... parts of the device should be handled as normal electronic waste at roll out Used batteries may also fall under the category of hazardous waste and should be disposed with proper care 15 Indication Contraindications Indications Suspected white coat hypertension Suspected nocturnal hypertension To establish dipper status Resistant hypertension Elderly patient As a guide to antihypertensive drug tr...

Страница 14: ...Humidity Non Condensing 10 95 Atmospheric Pressure 70 106 KPa Storage Transportation Temperature 20 to 50 C 4 to 122 F Humidity Non Condensing 10 95 Size 70 x 99 x 30 mm Weight Batteries Incl 240 g Blood Pressure Measurement Method Oscillometric Blood Pressure Maximum Storage Over 600 measurements Measurement Range Blood pressure 30 260 mmHg 4 35 kPa pulse 40 200 beat minute Passive Accuracy 3mmHg...

Страница 15: ...in the cuff or it is rather loose on the patient s arm E34 cuff not on arm The patient did not wear the cuff Device Error E90 device error The device could not measure due to a hardware error E99 device error The device does not start automatic measurement until the next programming 19 Product Warranty Information MONITOR WARRANTY The main monitor unit will be free from defects in materials and wo...

Страница 16: ...biosmedical com Please include your name return address phone number and email address BIOS Medical will repair or replace at BIOS Medical option free of charge any parts necessary to correct the defect in material or workmanship Please allow 10 days for repair and return shipping Always consult a physician for the interpretation of the blood pressure measurements Note that any blood pressure reco...

Страница 17: ...ion against environmental impact First digit 2 Protected against mid sized solid objects 12 mm Second digit 2 Protected against splash vertically in a max of 15 degrees This protection refers only to BP5 Serial number The first four digits of the serial number of a recorder show the year of production The rest is the serial number For example 2007 123456 This symbol shows that according to regulat...

Страница 18: ...low voltage power supply network which supplies buildings used for domestic use Harmonic emission IEC61000 3 2 Not applicable Voltage fluctuations Flickers IEC61000 3 3 Not applicable Electromagnetic immunity BIOS DiagnosticsTM ambulatory blood pressure monitor is suitable for use in the specified electromagnetic environment The purchaser or user of BIOS DiagnosticsTM products should assure that t...

Страница 19: ...P 01 Radiated RF IEC 61000 4 3 3V m 80MHz 2 5GHz 3V m Recommended Separation distance d 3 5 V1 P d 3 5 3V m P 80MHz 800MHz d 7 3V m P 800MHz 2 5GHz where P is the highest radiated power disclosed by the manufacturer of transmitter W and d is the recommended separation distance m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey1 a should be less than the co...

Страница 20: ...communication equipment transmitters and the device depending on the highest output power of the telecommunication equipment Separation distance in function of the frequency of the transmitter m The highest output power of the transmitter W 150KHz 80MHz d 3 5 V1 P 80MHz 800MHz d 3 5 E1 P 800MHz 2 5GHz d 7 E1 P 0 01 Not applicable 0 12 0 23 0 1 Not applicable 0 38 0 73 1 Not applicable 1 2 2 3 10 N...

Отзывы: