background image

 

 

USER MANUAL 

TD064-O&M-001 

REVISION 4  

 

 
 

Page 11 of 38 

 
 

 

 

 

3

 

HPV GENERATOR CONNECTION 

Note: The following instructions assume that the HPV Generator is 
located next to the chamber with a power outlet socket nearby. In other 
configurations the HPV Generator may be installed via fixed pipework in 
a technical space away from the chamber, or integrated into the 
chamber. 

Refer to the User Manual for the Generator with regard to loading Hydrogen 
Peroxide and setting up ready for operation. 

Connections to HPV Generator 

For HPV disinfection cycles, connections to the Generator can be via fixed 
installation pipework (if generator is to be located in a technical space away from 
the WIC) or by flexible hosing (as standard supplied with the Generator). 
Positioning of the WIC inlet / outlet connections will vary dependent upon Client 
requirements.  

 

 

 
For both fixed pipework and flexible 
hosing the connections to the 
Generator and WIC are Camlock 
connections. 
 
 
Camlock connector. 

The H

2

O

2

 return hose is fitted with a male Camlock connection to attach to the 

Generator and a female Camlock connection at the remote end to connect to the 
chamber. 

(Note: The colour coding of the hose and connector is blue). 

 

The H

2

0

2

 supply hose is fitted with a female Camlock connection to attach to the 

Generator and a male Camlock connection at the remote end to connect to the 
chamber.

 

(Note: This insulated hose goes to the red colour coded 

connection). 
 

The communication cable connection may be located in various positions on the 
chamber dependent upon Client requirements. 

 

There is a dedicated mains socket to power the HPV generator if the Generator is 
placed at the front of the WIC. If it is not located at the front then a dedicated 
power socket outlet must be within 2m of the HPV generator. 
 
 

Содержание WIC-1

Страница 1: ...R MANUAL TD064 O M 001 REVISION 4 Bioquell UK Ltd 52 Royce Close West Portway Andover Hampshire SP10 3TS T F E W 44 0 1264 835 835 44 0 1264 835 836 info bioquell com www bioquell com Page 1 of 38 Wal...

Страница 2: ......

Страница 3: ...ng Phase 10 2 3 3 Dwell Phase 10 2 3 4 Aeration Phase 10 3 HPV GENERATOR CONNECTION 11 Connections to HPV Generator 11 4 OPERATING INSTRUCTIONS 13 4 1 Loading the Chamber 13 4 2 Running a Cycle 14 4 2...

Страница 4: ...uled Maintenance 27 6 2 1 Parts to be replaced 28 6 2 2 Functional Checks 28 7 TROUBLESHOOTING 29 7 1 Alarms 29 7 2 Fault Finding 31 8 PARTS LIST 32 8 1 Replacement Parts 32 9 SPECIFICATION 33 10 EC D...

Страница 5: ...DROGEN PEROXIDE LIQUID can cause burns and blistering to the skin and tissue damage to the eyes when in contact HYDROGEN PEROXIDE VAPOUR can cause irritation to eyes nose throat lungs and skin and bre...

Страница 6: ...the HPV Generator and the WIC correctly The rings on the end of the Camlock levers ought to be fixed in the closed position with the hose connected BEFORE RUNNING A CYCLE Ensure all personnel are out...

Страница 7: ...SOPs Repairs or adjustments to the equipment should only be made by trained Bioquell personnel Maintenance performed by unqualified personnel or installation of unauthorised parts could invalidate the...

Страница 8: ...ther side door can be opened Each time a door is closed clean up will be completed automatically before re entry door opening is possible The Hydrogen Peroxide Vapour disinfection cycle is an automate...

Страница 9: ...require a dedicated extract 2 2 5HVAC If the facility has an HVAC system connected the chamber will be supplied with inlet and outlet connection points 2 2 6Ceiling Mounted Aeration Unit Optional aer...

Страница 10: ...is a closed loop process the returning air vapour mixture from the Chamber continues to pass through the gassing circuit for a pre programmed period At the end of the pre set time the cycle automatica...

Страница 11: ...ctions will vary dependent upon Client requirements For both fixed pipework and flexible hosing the connections to the Generator and WIC are Camlock connections Camlock connector The H2O2 return hose...

Страница 12: ...Connect hoses to chamber as shown in photo Connect the communications cable to the chamber Secured Camlock arms Note It is recommended that the Camlock arms once secured to the generator and chamber...

Страница 13: ...hat Door A has priority i e loading from Door A Whilst the pneumatic seals are releasing the display will show DOOR UNLOCKING Do not try to force the door open Open door If required the light of the c...

Страница 14: ...g during the cycle such as ventilation units and electronic equipment to ensure good gas distribution within them ought to be plugged in to the power sockets Close the door The above screens would be...

Страница 15: ...operational in accordance with the instructions within the specific Generator User Manual When about to start a cycle check that all the connections have been made correctly Check the hose carrying v...

Страница 16: ...en Peroxide The Bioquell L 2 does not have this functionality and as such the Operator must ensure enough Peroxide is available in the bottle Bioquell L 2 Bioquell IG 1 L 3 The cycle can be selected b...

Страница 17: ...s button to return to main menu or button to view status of chamber in OFF mode The screen shown on the left shows insufficient Hydrogen Peroxide to run the selected cycle equally the setup prompt may...

Страница 18: ...ambient conditions within chamber in OFF mode To start cycle ensure generator is connected properly press button and confirm connection complete Bioquell L 2 Bioquell IG 1 L 3 In either case To initia...

Страница 19: ...itioning phase of cycle Press WIC TREND to display ambient conditions within chamber This screen displays the chamber during the gassing phase If the cycle is aborted once in the gassing phase the sys...

Страница 20: ...f Aeration the display will prompt the Operator to check the Hydrogen Peroxide concentration in the chamber Press OK The concentration can be checked from either side using an independent low level se...

Страница 21: ...age the Operator can proceed to unlock the door and remove the contents 4 3 Unloading the Chamber Unlock the clean side door Remove the items in the Chamber and place any items which require passing b...

Страница 22: ...f by the following method Turn OFF and disconnect power to the Generator Disconnect the hoses and control cable Place the Generator into the chamber suitably configure it and connect to power supply a...

Страница 23: ...TOR level or above password required to access 5 1 WIC Accessible from the MAIN MENU this screen displays the status of the chamber From here the Operator can turn on the ceiling light view the ambien...

Страница 24: ...is only viewable and displays the chamber Serial Number software versions and calibration due date 5 3 2Manage Passwords User s name is displayed here By touching here the password of the person logg...

Страница 25: ...ting up new users re setting passwords and assigning user groups to the user ENGINEERS As per ADMINISTRATORS but also has access to the System Engineering screen as well as able to edit the machine se...

Страница 26: ...ply follow the instructions and it will ask for a series of crosses to be touched in various places on the screen Clean Screen When pressed disables any actions initiated by touching it for a limited...

Страница 27: ...hat no adjustment of the screen is done whilst cleaning Navigate to this button via Main menu System Admin Screen Management Password access ADMINISTRATOR or above is required to enter the System Admi...

Страница 28: ...any broken or damaged parts Replacement of H2O2 sensor head with a new pre calibrated head Replacement of RH Sensor and Temperature Sensor with pre calibrated sensors Replacement of Carbon and HEPA fi...

Страница 29: ...gency Aeration Alarm No Alarm Message Condition Action User Action 47 DOOR A SEAL FAULT If Door A should be locked and door seal pressure less than limit or if door should be unlocked and pressure gre...

Страница 30: ...n limit If in cycle goes to OFF mode Check door seal for damage Attempt to re lock door If OK contact Service Provider 61 DOOR B UNLOCKED The door is open when it is not enabled If in cycle goes to OF...

Страница 31: ...vestigate BMS system and its signals If OK contact BMS supplier 78 BMS FAULT HVAC OPTION 1 disabled If time expired and signal not received If in cycle goes to OFF mode Investigate BMS system and its...

Страница 32: ...sembly Single Door Solenoid H19020500 Emergency Override H05990506 HEPA Filter H14 305 x 305 x 110mm H03990514 HI1 H2O2 Sensor Head H19010004 M FU1 FU2 2A Quick Blow Fuse 5 x 20mm H06010508 FU3 FU4 4A...

Страница 33: ...supply Internal Dimensions WxHxD 1680 or 2180 x 2200 x 2320 or 2720 3120 mm 66 1 or 85 8 x 86 6 x 91 3 or 107 123 inches Doorway Opening Size WxH 1100 x 2200 mm 43 3 x 86 6 inches Control System Siem...

Страница 34: ...USER MANUAL TD064 O M 001 REVISION 4 Page 34 of 38 10 EC DECLARATION OF CONFORMITY...

Страница 35: ...r unlocking To start cycle either navigate to main menu and select HPV CYCLE or start from HPV generator Level 2 Level 4 Level 3 Level 1 Only displayed if HMI Clarus Start enabled Only displayed if HP...

Страница 36: ...USER MANUAL TD064 O M 001 REVISION 4 Page 36 of 38 Screen Management Manage Passwords System Info Time Date Password access ADMINISTRATORS password level required to enter SYSTEM ADMIN menu...

Страница 37: ...viders 5 The warranty is conditional on routine maintenance being carried out in accordance with the maintenance schedule described in the User Manual 6 Only fully trained and Bioquell authorised Serv...

Страница 38: ...REVISION 4 Page 38 of 38 For further information on Bioquell s technology Bioquell UK Ltd 52 Royce Close West Portway Andover Hampshire SP10 3TS T 44 0 1264 835 835 F 44 0 1264 835 836 E info bioquel...