BSLSTM SPECIFICATIONS

 

(This older unit uses dial reading and a flip range switch)

  

Pulse width

 

 

Controlled by: 

Computer, with lockable width limit 

Range: 

50 microseconds 

– 100 milliseconds 

Resolution: 

2 microseconds 

Accuracy: 

5% (Can be improved to better than 2% using the “Correction factor” in the 
“Stimulator Preferences’ window.) 

Correction factor 

Range: 0 - 150 microseconds 
Average value: 110 microseconds 

Pulse Repetition

 

 

Controlled by: 

Computer-based software 

Pattern: 

Selectable (1-254 pulses) or continuous 

Range

—No Load: 

5 seconds - .3 milliseconds Period (.2 - 3,333 Hz Rate) 

Range

—Load: 

2 K Ohm load 
0 - 10 Volt Range:  5 seconds to the following minimum repetition period: 
 

100 ms P.W. 

150 ms 

 

10 ms P.W. 

10.1 ms 

 

1 ms P.W. 

1.1 ms 

0 - 100 Volt Range:  5 seconds to the following minimum repetition period: 
 

100 ms P.W.  

100 Volts: 

 beyond functional limits 

 

 

50 Volts:  

250 ms 

 

10 ms P.W. 

100 Volts: 

200 ms 

 

 

50 Volts:  

150 ms 

 

1 ms P.W. 

100 Volts: 

20 ms 

 

 

50 Volts:  

2.5 ms 

Limits: 

User adjustable lower and upper rate limits 

Resolution: 

2 microseconds 

Accuracy: 

Better than 2% 

Initial Pulse Delay

 

 

Time range: 

None or .5 - 100 milliseconds 

Resolution: 

2 microseconds 

Pulse level

 

 

Controlled by: 

Manually (10 turn potentiometer) 

Range (switchable): 

Range 1

: .025 - 10 Volts 

Range 2

: .15 - 100 Volts 

Infinite (potentiometer adjustable) range 

Accuracy: 

5% accuracy to dial indicator 

Reference Output

 

Correlates to actual pulse output (Requires Calibration) 

Pulse width: 

15 millisecond fixed pulse width 

Amplitude: 

0 - 10 mV correlates to 0 

– 10 V actual output or 0 – 100 V actual output 

Manual Test Pulse

 

(Button on back panel) 

Note

: Will only function when “Trigger” cable is not connected to the MP3X. 

Pulse Width:

 

2.5 - 3 milliseconds 

Stimulator isolation 

 

Volts: 

2,000 Volts DC (HI POT test) 

Capacitance 
coupling: 

60 pF

 

Power requirements

 

12 Volts DC adapter (included), 1 Amp 

Fuse 

 

250 V, 2 A, fast blow

 

Dimensions: 

1.25” length 

 

.25” diameter 

Module Weight

 

610 grams 

Module Dimensions

 

16 cm x 16 cm x 5 cm 

 

Содержание BSLSTMA

Страница 1: ...ng and a flip range switch The same guidelines and cautions described here apply except when noted The BSLSTM Stimulator works in conjunction with the Biopac Student Lab System to allow precise stimul...

Страница 2: ...menu in order to maintain direct Level recordings If the Range is changed during recording the user should manually enter a software marker to note the change by holding down F9 on a PC or Esc key on...

Страница 3: ...cquisition unit The MP3X sends the Pulse width and Pulse rate information via this cable Manual Test button Used to diagnose problems with the BSLSTM stimulator unit When the Trigger and Reference Out...

Страница 4: ...obtain the signal representing 0 V out of the stimulator 5 Add the Input value found in Cal 2 to the Input Value displayed for Cal 1 For example if Cal 2 is pressed and provides an Input Value of 255...

Страница 5: ...r from the back of the BSLSTM into an open channel input port DB9 female connectors CH 1 CH 2 CH 3 or CH 4 on the front of the MP3X 7 Plug the 12 Volt DC adapter into the wall 8 Mate the DC output con...

Страница 6: ...P W 100 Volts 400 ms 50 Volts 30 ms 1 ms P W 100 Volts 4 ms 50 Volts 3 ms Limits User adjustable lower and upper rate limits Resolution 2 microseconds Accuracy Better than 2 Initial Pulse Delay Time...

Страница 7: ...Volts beyond functional limits 50 Volts 250 ms 10 ms P W 100 Volts 200 ms 50 Volts 150 ms 1 ms P W 100 Volts 20 ms 50 Volts 2 5 ms Limits User adjustable lower and upper rate limits Resolution 2 micr...

Страница 8: ...lead with a BNC connector such as needle electrodes or clip leads for direct stimulation of animal or tissue preps Control the stimulus with the Output Control option of the BSL PRO software Output c...

Страница 9: ...uld only be used under direct supervision of an Instructor Never place any stimulator leads in the mouth or any other body orifice To prevent a Ground loop the Ground of the stimulator electrode and t...

Отзывы: