As shown in the following diagram, the indicator mark is between the two dials. 

 

 

     

Close-up of “Level” adjustment knob 

Stimulus output

 

Stimulus pulse output for connection to external electrodes or other devices. This is a 
standard BNC style connector. 

Pulse indicator

 

LED flashes when the stimulus pulse is active: BSLSTMA/B = red. BSLSTM = green. 

Power indicator

 

Activated when the DC adapter is plugged in and the power switch on the back panel is 
turned ON. 
BSLSTMA/B:  The LCD display is activated. 
BSLSTM:  LED indicator lights green  

BACK PANEL TERMINOLOGY 

 

Power switch 

Rocker switch for turning the BSLSTM power ON and OFF. 

Fuse holder 

If the fuse blows and must be replaced, use a screwdriver to open (counterclockwise) and 
close (clockwise) the fuse cap. 

DC Input 

Socket for BIOPAC DC adapter. 

Trigger cable

 

Connects to the Analog Out connector on the back of the MP3X acquisition unit. The 
MP3X sends the Pulse width and Pulse rate information via this cable. 
 

Manual Test 
button

 

Used to diagnose problems with the BSLSTM stimulator unit. 

When the 

Trigger

 and 

Reference

 

Output

 cables are 

disconnected

 from 

the MP3X , the 

Manual Test

 button can be used to initiate a stimulus with 

a fixed pulse width of 2.5 milliseconds. 

 

Reference Output 
Cable 

 
 
 
 
 

 

The stimulus marker output is labeled 

Reference Output

 on the back panel of the 

BSLSTM. This output cable connects to any of the four channel inputs (CH1, CH 2, CH 
3, or CH 4) on the front of the MP3X acquisition unit. The output cable carries the 
stimulator marker pulse to the MP3X. The marker pulse has a fixed pulse width 15 ms 
and is generated each time the stimulator generates a pulse. 

 

BSLSTMA/B: Use the front panel Reference switch to select Actual or Fixed.

 

 

BSLSTM has a fixed pulse width of 15 ms, selected so that the MP30 can capture 
the pulse with a sample rate as low as 100 samples per second.

 

If the BSL 

PRO

 software has been setup correctly, the amplitude of this marker will 

reflect the 

Level

 knob setting on the BSLSTM. See the 

Range switch

 section for 

information on how this reading can be affected. 

 
 

Содержание BSLSTMA

Страница 1: ...ng and a flip range switch The same guidelines and cautions described here apply except when noted The BSLSTM Stimulator works in conjunction with the Biopac Student Lab System to allow precise stimul...

Страница 2: ...menu in order to maintain direct Level recordings If the Range is changed during recording the user should manually enter a software marker to note the change by holding down F9 on a PC or Esc key on...

Страница 3: ...cquisition unit The MP3X sends the Pulse width and Pulse rate information via this cable Manual Test button Used to diagnose problems with the BSLSTM stimulator unit When the Trigger and Reference Out...

Страница 4: ...obtain the signal representing 0 V out of the stimulator 5 Add the Input value found in Cal 2 to the Input Value displayed for Cal 1 For example if Cal 2 is pressed and provides an Input Value of 255...

Страница 5: ...r from the back of the BSLSTM into an open channel input port DB9 female connectors CH 1 CH 2 CH 3 or CH 4 on the front of the MP3X 7 Plug the 12 Volt DC adapter into the wall 8 Mate the DC output con...

Страница 6: ...P W 100 Volts 400 ms 50 Volts 30 ms 1 ms P W 100 Volts 4 ms 50 Volts 3 ms Limits User adjustable lower and upper rate limits Resolution 2 microseconds Accuracy Better than 2 Initial Pulse Delay Time...

Страница 7: ...Volts beyond functional limits 50 Volts 250 ms 10 ms P W 100 Volts 200 ms 50 Volts 150 ms 1 ms P W 100 Volts 20 ms 50 Volts 2 5 ms Limits User adjustable lower and upper rate limits Resolution 2 micr...

Страница 8: ...lead with a BNC connector such as needle electrodes or clip leads for direct stimulation of animal or tissue preps Control the stimulus with the Output Control option of the BSL PRO software Output c...

Страница 9: ...uld only be used under direct supervision of an Instructor Never place any stimulator leads in the mouth or any other body orifice To prevent a Ground loop the Ground of the stimulator electrode and t...

Отзывы: