BIOPAC Systems BSLSTMA Скачать руководство пользователя страница 5

 

 

 

PRODUCT SHEET 

[email protected] 

[email protected]

 

www.biopac.com 

 

 

BIOPAC Hardware   |   

STIMULATORS MP3X

   |   Page 5 - 9 

Updated: 9.12.2019

 

CONNECTING THE BSLSTM TO THE MP3X  

1)

 

Turn the 

MP3X 

unit 

OFF

2)

 

Confirm that 

Power

 switch on the back of the 

BSLSTM

 is in the 

OFF

 position. 

3)

 

Set the 

Range

 on the front of the 

BSLSTM

 to 

0-10 V

4)

 

Set the 

Level

 to 1 Volt. 

 

BSLSTM: 1 Volt is set when the Major Scale (top number) is 1 and the Minor Scale (lower number) 
is 0. 

5)

 

Plug the 

Trigger

 cable (female DB9 connector) from the back of the 

BSLSTM

 into the 

Analog

 

Out

 port 

(DB9 Male connector) on the back of the 

MP3X

 

6)

 

Plug the 

Reference Output

 cable (Male DB9 connector) from the back of the 

BSLSTM

 into an open 

channel input port (DB9 female connectors: CH 1, CH 2, CH 3, or CH 4) on the front of the 

MP3X

7)

 

Plug the 12 Volt 

DC adapter

 into the wall. 

8)

 

Mate the 

DC output

 connector on the end of the adapter cable to the 

DC Input

 socket on the back of the 

BSLSTM

 

Make sure the connector is pressed in completely. 

9)

 

Plug the stimulator electrode assembly into the BNC connector on the front of the stimulator, labeled 
Output on the BSLSTMA/B and Stimulus Output on the BSLSTM. 

10)

 

Place the BSLSTMA/B unit away from the MP3X. Placing the BSLSTMA/B too close to MP3X 
hardware can result in data distortion of the BSLSTMA/B pulse width signal; the distortion is more 
apparent at higher sampling rates. 

 

NEVER set the BSLSTMA/B atop an MP3X. 

 

Position the BSLSTMA/B away from the MP3X to reduce the signal distortion. 

 

Содержание BSLSTMA

Страница 1: ...ng and a flip range switch The same guidelines and cautions described here apply except when noted The BSLSTM Stimulator works in conjunction with the Biopac Student Lab System to allow precise stimul...

Страница 2: ...menu in order to maintain direct Level recordings If the Range is changed during recording the user should manually enter a software marker to note the change by holding down F9 on a PC or Esc key on...

Страница 3: ...cquisition unit The MP3X sends the Pulse width and Pulse rate information via this cable Manual Test button Used to diagnose problems with the BSLSTM stimulator unit When the Trigger and Reference Out...

Страница 4: ...obtain the signal representing 0 V out of the stimulator 5 Add the Input value found in Cal 2 to the Input Value displayed for Cal 1 For example if Cal 2 is pressed and provides an Input Value of 255...

Страница 5: ...r from the back of the BSLSTM into an open channel input port DB9 female connectors CH 1 CH 2 CH 3 or CH 4 on the front of the MP3X 7 Plug the 12 Volt DC adapter into the wall 8 Mate the DC output con...

Страница 6: ...P W 100 Volts 400 ms 50 Volts 30 ms 1 ms P W 100 Volts 4 ms 50 Volts 3 ms Limits User adjustable lower and upper rate limits Resolution 2 microseconds Accuracy Better than 2 Initial Pulse Delay Time...

Страница 7: ...Volts beyond functional limits 50 Volts 250 ms 10 ms P W 100 Volts 200 ms 50 Volts 150 ms 1 ms P W 100 Volts 20 ms 50 Volts 2 5 ms Limits User adjustable lower and upper rate limits Resolution 2 micr...

Страница 8: ...lead with a BNC connector such as needle electrodes or clip leads for direct stimulation of animal or tissue preps Control the stimulus with the Output Control option of the BSL PRO software Output c...

Страница 9: ...uld only be used under direct supervision of an Instructor Never place any stimulator leads in the mouth or any other body orifice To prevent a Ground loop the Ground of the stimulator electrode and t...

Отзывы: