CONNECTING THE BSLSTM TO THE MP3X
1)
Turn the
MP3X
unit
OFF
.
2)
Confirm that
Power
switch on the back of the
BSLSTM
is in the
OFF
position.
3)
Set the
Range
on the front of the
BSLSTM
to
0-10 V
.
4)
Set the
Level
to 1 Volt.
➢
BSLSTM: 1 Volt is set when the Major Scale (top number) is 1 and the Minor Scale (lower number)
is 0.
5)
Plug the
Trigger
cable (female DB9 connector) from the back of the
BSLSTM
into the
Analog
Out
port
(DB9 Male connector) on the back of the
MP3X
.
6)
Plug the
Reference Output
cable (Male DB9 connector) from the back of the
BSLSTM
into an open
channel input port (DB9 female connectors: CH 1, CH 2, CH 3, or CH 4) on the front of the
MP3X
.
7)
Plug the 12 Volt
DC adapter
into the wall.
8)
Mate the
DC output
connector on the end of the adapter cable to the
DC Input
socket on the back of the
BSLSTM
.
➢
Make sure the connector is pressed in completely.
9)
Plug the stimulator electrode assembly into the BNC connector on the front of the stimulator, labeled
Output on the BSLSTMA/B and Stimulus Output on the BSLSTM.
10)
Place the BSLSTMA/B unit away from the MP3X. Placing the BSLSTMA/B too close to MP3X
hardware can result in data distortion of the BSLSTMA/B pulse width signal; the distortion is more
apparent at higher sampling rates.
•
NEVER set the BSLSTMA/B atop an MP3X.
•
Position the BSLSTMA/B away from the MP3X to reduce the signal distortion.