- Avoid contact with dripping or splashing liquids.
- The minimum blood sample size to test using the

 

 

Blood Glucose Monitoring System 

 is 0.75µL: (     )

TM

Rightest

GM700S

5

4

 

Caution

Caution

 

Blood sample size above 3.0µL might contaminate the test strip port and the meter.
Blood sample size below 0.75µL may cause an inaccurate result or may prevent a meter reading. 
An Er4 reading will be displayed if the sample size is too small. In this case, repeat the test with a 
new test strip.

Sample Size Example

Important Safety Notes:

- Users should wash their hands thoroughly with soap and water before and after handling 

the meter, lancing device, or test strips.

0.75µL       1.0µL        1.5µL            2.0µL           3.0µL 

- Before using the

 

 

Blood Glucose Monitoring System 

to test your blood 

glucose, please read all of the instructions and conduct all of the tests including the quality control 
test

- Please perform the quality control test regularly to make sure the test results are accurate.

TM

TM

Rightest

 

GM700S  

Rightest

GS700 Other brands’ test strips should not be used under any circumstances. The 

use of other brands’ strips may give inaccurate results.

TM

 The

 

Rightest

 Blood Glucose Monitoring System GM700S is intended for in vitro diagnostic use 

TM

only. The

 

Rightest

 Blood Glucose test results using fresh capillary whole blood samples from 

the fingertip, palm and forearm are calibrated to be the equivalent to that of plasma testing. 
Venous, neonatal or arterial blood may also be used only if drawn by healthcare professionals.

TM

Rightest

GM700S

TM

- If the

 

Rightest

Blood Glucose Meter GM700S and Test Strips GS700 are exposed to a 

substantial change in temperature, please wait 30 minutes before measurement.

- Follow all environmental protection regulations when disposing of batteries.

TM

Rightest

GM700S 

 (Refer to page 37). 

- The

 

Blood Glucose Meter 

can only be used with

 

 Blood Glucose 

Test Strips 

-

- The

 

 Blood Glucose Monitoring System 

 is intended for self-testing. It should 

not be used to diagnose diabetes mellitus.

 

The system has been tested with neonatal blood. As a matter of good clinical practice, caution is 
advised in the interpretation of neonate glucose value below 50 mg/dL. Please follow the 
recommendations for follow-up care that have been set by your intuition for critical glucose values 
in neonates. Results for neonates exhibiting symptoms of galactosemia should be confirmed by 
laboratory tests.

Содержание rightest GM700S

Страница 1: ...Berneck Switzerland E mail info bionime ch GM700S 0197 BIONIME CORPORATION No 100 Sec 2 Daqing St South Dist Taichung City 40242 Taiwan Tel 886 4 2369 2388 Fax 886 4 2261 7586 http www bionime com E...

Страница 2: ......

Страница 3: ...by Bionime Corporation If you have any questions or concerns please contact your local Customer Service or call us at 886 0800 371 688 GMT 08 00 08 30 17 30 Bionime Customer Thank you for selecting th...

Страница 4: ...ay give inaccurate results TM The Rightest Blood Glucose Monitoring System GM700S is intended for in vitro diagnostic use TM only The Rightest Blood Glucose test results using fresh capillary whole bl...

Страница 5: ...Alternative Site Testing View Window Appearance Understanding Test Results and Messages TM Rightest GM700S TM Rightest GM700S TM Rightest GM700S TM Rightest GS700 TM Rightest GS700 8 10 14 16 18 22 2...

Страница 6: ...ucose Meter with 1 CR2032 batteries installed 6 Blood Glucose Test strips GS700 10 25 pcs 7 Control Solution GC700 Normal Level is provided in your starter kit 8 Lancing Device 9 Clear Cap 10 Disposab...

Страница 7: ...The Blood Glucose Meter TM Rightest GM700S Right Button Battery Cover Meter Battery The Blood Glucose Meter TM Rightest GM700S Left Button Main Button Display Window Test Strip Port 2D barcode with s...

Страница 8: ...esult Test result Warns when the batteries are low or must be replaced Indicates when to apply the blood sample Appears after you insert test strip into meter Appears when an error occurs Current date...

Страница 9: ...est Strip GS700 record the date on the vial TM Rightest Blood Glucose Test Strip GS700 TM If the Rightest Blood Glucose Meter and Test Strips are exposed to a substantial change in temperature please...

Страница 10: ...ease follow the local regulation to properly discard a used battery 4 The Blood Glucose Meter 5 Press any button to exit the self test and enter Setting Mode TM Rightest replacing the battery all symb...

Страница 11: ...onds During this time the screen will go blank until you hear a beep After the beep the meter will turn on into the Setting Mode The display screen will show setting data GM700S GM700S Setting Up Your...

Страница 12: ...change any settings during Meter Set up for over 2 minutes the Blood Glucose Meter will leave setting mode and power off automatically TM Rightest GM700S Setting Up Your Meter Setting the Date Time a...

Страница 13: ...no strip is inserted GM700S GM700S GM700S Turning On Off the Meter Handling the Blood Glucose Test Strip TM Rightest GS700 How to handle the ood Glucose Test Strip Inserting the Blood Glucose Test St...

Страница 14: ...f the test strip port Please follow your local regulations and discard used strips properly TM Rightest GS700 1 2 3 Getting Ready for Testing Before performing a blood glucose test prepare the items b...

Страница 15: ...ed skin 1 2 3 4 5 6 Performing a Blood Glucose Test 7 Hold the base in one hand and pull on the plunger with the other hand The device will be cocked Release the plunger and it will automatically move...

Страница 16: ...repeat with a new Blood Glucose Test Strip GS700 15 Countdown Mode will begin on the display window After 5 seconds your test result will appear TM Rightest 13 15 Performing a Blood Glucose Test 14b 1...

Страница 17: ...uld wash hands thoroughly with soap and water after handling the meter lancing device and test strips TM Rightest TM TM Rightest Rightest GM700S Setting Markers According to actual life status you can...

Страница 18: ...ratory test system that is known to be well maintained and monitored by a healthcare provider use the clear TM cap provided with the Rightest Lancing Device GD500 when testing Alternative Site Testing...

Страница 19: ...provider for appropriate disposal of used test strips and lancets TM Rightest TM Rightest TM TM Rightest Rightest GM700S Make sure your blood sample covers the whole area of the view window to get an...

Страница 20: ...t TM Rightest Control Solution Range Example of Control Solution Range printed on your test strip vial label Control Solution Range mg dL Low 28 48 Normal 107 145 High 315 426 NOTE Normal of Control S...

Страница 21: ...ate 2014 06 When should a Quality Control Test Be Performed To ensure that your Blood Glucose Meter and Blood Glucose Test Strip are working properly To confirm that you are following the correct test...

Страница 22: ...trol solution This could contaminate the meter via the test strip port Keep the test strip port clean and dry Clean immediately if the test strip port is stained or is overly exposed to moisture Do no...

Страница 23: ...lts TM Rightest Possible reasons your Control Solution results are out of the range Your Control solution GC700 is expired or was first opened more than 3 months ago Your Blood Glucose Test Strip GS70...

Страница 24: ...lts control solution test results and test results under abnormal temperature 43 F 6 C 111 F 44 C will be excluded from average calculations The Average function requires that the correct time and dat...

Страница 25: ...ort is stained with blood control solution or any liquid please use a dry tissue or alcohol swab to clean it immediately Do not use anything wet to clean it Perform a quality control test to ensure th...

Страница 26: ...The meter should work normally after finishing the above steps If the meter still does not work please contact your local customer service TM Rightest GM700S Bionime Strip Error TM Rightest TM Rightes...

Страница 27: ...other interference keep meter out of electromagnetic radiation sources TM Rightest Inaccurate test results may be obtained at altitudes greater than 10 000 feet 3048 meters above sea level Xylose 20 m...

Страница 28: ...m x 27 5 mm 14 140 F 10 60 C 90 relative humidity 39 86 F 4 30 C Measurement Technology Sample Minimum Sample Volume Measuring Range Test Time Memory Capacity Power Saving Operating Temperature Operat...

Страница 29: ...ghtest GM700S TM Different models have different specifications This warranty applies only to the Rightest Blood Glucose Monitoring System GM700S other models are not covered with this warranty card T...

Страница 30: ...Taichung City 40242 Taiwan Product complies with Medical Device Directive 93 42 EEC EC Rep Disposable Sterile Lancets Manufacturer SteriLance Medical SuZhou Inc No 68 LiTangHe RD XiangCheng SuZhou Ji...

Отзывы: