37

36

Understanding Test Results and Messages

About Quality Control Testing

What is a Quality Control Test ?
To ensure proper monitoring function, it is necessary to regularly perform a quality control test.
Use one of the 

 Control Solutions GC700 when testing your 

 Blood Glucose 

Monitoring System GM700S in the Control Solution Mode. If the test result is within the Control 
Solution Range printed on the strip vial label, the 

Blood Glucose Monitoring System 

GM700S passes the Quality Control Test and your 

 Blood Glucose Monitoring System 

GM700S is working properly.

TM

TM

Rightest

Rightest

TM

Rightest

TM

Rightest

 

Control Solution Range:

Example of Control Solution Range 
printed on your test strip vial label.

Control Solution Range mg/dL

Low:                 28-48

Normal:          107-145
High:              315-426  

NOTE

- Normal of 

 Control Solution GC700 is provided in your starter kit. If you want to 

purchase, High or another Normal control solutions, please contact customer service.

TM

 

Rightest

The 

 Meter

 displays results between 10 

and 600 mg/dL.

If the test result is below 10 mg/dL , “      ” will appear on the screen. Please 
repeat your test with by a new test strip.

If you still get a “      ” result, contact your healthcare professional.

If the test result is above 600 mg/dL, “      ” will appear on the screen. Please 
repeat your test again with a new test strip. 

If you still get a “      ” result, contact your healthcare professional.

TM

 

Rightest

 Blood Glucose

 GM700S

Содержание rightest GM700S

Страница 1: ...Berneck Switzerland E mail info bionime ch GM700S 0197 BIONIME CORPORATION No 100 Sec 2 Daqing St South Dist Taichung City 40242 Taiwan Tel 886 4 2369 2388 Fax 886 4 2261 7586 http www bionime com E...

Страница 2: ......

Страница 3: ...by Bionime Corporation If you have any questions or concerns please contact your local Customer Service or call us at 886 0800 371 688 GMT 08 00 08 30 17 30 Bionime Customer Thank you for selecting th...

Страница 4: ...ay give inaccurate results TM The Rightest Blood Glucose Monitoring System GM700S is intended for in vitro diagnostic use TM only The Rightest Blood Glucose test results using fresh capillary whole bl...

Страница 5: ...Alternative Site Testing View Window Appearance Understanding Test Results and Messages TM Rightest GM700S TM Rightest GM700S TM Rightest GM700S TM Rightest GS700 TM Rightest GS700 8 10 14 16 18 22 2...

Страница 6: ...ucose Meter with 1 CR2032 batteries installed 6 Blood Glucose Test strips GS700 10 25 pcs 7 Control Solution GC700 Normal Level is provided in your starter kit 8 Lancing Device 9 Clear Cap 10 Disposab...

Страница 7: ...The Blood Glucose Meter TM Rightest GM700S Right Button Battery Cover Meter Battery The Blood Glucose Meter TM Rightest GM700S Left Button Main Button Display Window Test Strip Port 2D barcode with s...

Страница 8: ...esult Test result Warns when the batteries are low or must be replaced Indicates when to apply the blood sample Appears after you insert test strip into meter Appears when an error occurs Current date...

Страница 9: ...est Strip GS700 record the date on the vial TM Rightest Blood Glucose Test Strip GS700 TM If the Rightest Blood Glucose Meter and Test Strips are exposed to a substantial change in temperature please...

Страница 10: ...ease follow the local regulation to properly discard a used battery 4 The Blood Glucose Meter 5 Press any button to exit the self test and enter Setting Mode TM Rightest replacing the battery all symb...

Страница 11: ...onds During this time the screen will go blank until you hear a beep After the beep the meter will turn on into the Setting Mode The display screen will show setting data GM700S GM700S Setting Up Your...

Страница 12: ...change any settings during Meter Set up for over 2 minutes the Blood Glucose Meter will leave setting mode and power off automatically TM Rightest GM700S Setting Up Your Meter Setting the Date Time a...

Страница 13: ...no strip is inserted GM700S GM700S GM700S Turning On Off the Meter Handling the Blood Glucose Test Strip TM Rightest GS700 How to handle the ood Glucose Test Strip Inserting the Blood Glucose Test St...

Страница 14: ...f the test strip port Please follow your local regulations and discard used strips properly TM Rightest GS700 1 2 3 Getting Ready for Testing Before performing a blood glucose test prepare the items b...

Страница 15: ...ed skin 1 2 3 4 5 6 Performing a Blood Glucose Test 7 Hold the base in one hand and pull on the plunger with the other hand The device will be cocked Release the plunger and it will automatically move...

Страница 16: ...repeat with a new Blood Glucose Test Strip GS700 15 Countdown Mode will begin on the display window After 5 seconds your test result will appear TM Rightest 13 15 Performing a Blood Glucose Test 14b 1...

Страница 17: ...uld wash hands thoroughly with soap and water after handling the meter lancing device and test strips TM Rightest TM TM Rightest Rightest GM700S Setting Markers According to actual life status you can...

Страница 18: ...ratory test system that is known to be well maintained and monitored by a healthcare provider use the clear TM cap provided with the Rightest Lancing Device GD500 when testing Alternative Site Testing...

Страница 19: ...provider for appropriate disposal of used test strips and lancets TM Rightest TM Rightest TM TM Rightest Rightest GM700S Make sure your blood sample covers the whole area of the view window to get an...

Страница 20: ...t TM Rightest Control Solution Range Example of Control Solution Range printed on your test strip vial label Control Solution Range mg dL Low 28 48 Normal 107 145 High 315 426 NOTE Normal of Control S...

Страница 21: ...ate 2014 06 When should a Quality Control Test Be Performed To ensure that your Blood Glucose Meter and Blood Glucose Test Strip are working properly To confirm that you are following the correct test...

Страница 22: ...trol solution This could contaminate the meter via the test strip port Keep the test strip port clean and dry Clean immediately if the test strip port is stained or is overly exposed to moisture Do no...

Страница 23: ...lts TM Rightest Possible reasons your Control Solution results are out of the range Your Control solution GC700 is expired or was first opened more than 3 months ago Your Blood Glucose Test Strip GS70...

Страница 24: ...lts control solution test results and test results under abnormal temperature 43 F 6 C 111 F 44 C will be excluded from average calculations The Average function requires that the correct time and dat...

Страница 25: ...ort is stained with blood control solution or any liquid please use a dry tissue or alcohol swab to clean it immediately Do not use anything wet to clean it Perform a quality control test to ensure th...

Страница 26: ...The meter should work normally after finishing the above steps If the meter still does not work please contact your local customer service TM Rightest GM700S Bionime Strip Error TM Rightest TM Rightes...

Страница 27: ...other interference keep meter out of electromagnetic radiation sources TM Rightest Inaccurate test results may be obtained at altitudes greater than 10 000 feet 3048 meters above sea level Xylose 20 m...

Страница 28: ...m x 27 5 mm 14 140 F 10 60 C 90 relative humidity 39 86 F 4 30 C Measurement Technology Sample Minimum Sample Volume Measuring Range Test Time Memory Capacity Power Saving Operating Temperature Operat...

Страница 29: ...ghtest GM700S TM Different models have different specifications This warranty applies only to the Rightest Blood Glucose Monitoring System GM700S other models are not covered with this warranty card T...

Страница 30: ...Taichung City 40242 Taiwan Product complies with Medical Device Directive 93 42 EEC EC Rep Disposable Sterile Lancets Manufacturer SteriLance Medical SuZhou Inc No 68 LiTangHe RD XiangCheng SuZhou Ji...

Отзывы: