Before Requesting After-Sales Service
35
BCM-600
Before Requesting After-Sales Service
This section shows how to treat simple problems encountered during use.
Please check the following points before requesting after-sales service
▣
Problems in Display
Sign
How to treat
▣
Problems in Measurement ECG
Sign
How to treat
The ECG waveform is not
displayed.
Electrocardiograph (Ver 1.0)
34
Before Requesting After-Sales Service
▣
Problems in Printer (OPTION)
Sign
How to treat
1. Press the power button during 1~2 seconds
2. Check if “Power off” message is shown in display.
1. Check if there is any equipment with a strong magnetism
around the product.
2. Check if the cable connection is stable.
3. Check if the product is grounded.
The power of the product is
not turned off.
The waveform is not normal.
The power of the product
turns off suddenly during use.
The power of the product is
not turned on.
1. Check if the AC power is supplied..
2. Press and hold the power button for 2 seconds
1. Check if the power is off due to battery release.
2. Check if AC power is supplied to the product.
1. Check if the electrode attached to the patient is stable.
1. Check if the cable connection is stable.
2. Check if the ECG cable is damaged.
Severe noise on the
waveform.
No printing
1. Check if the printer icon blinks on the screen.
2. Check if the paper is set normally.
3. Check if the printer door is closed.
Noise during printing
1. Check if the roller is contaminated with foreign materials
or if the gear is out of position.
2. Check if the problem is due to a paper jam (change the
paper)