When the Background Calibration is complete, click
Close.
NOTE:
The door will open when the Background Calibration is successfully complete. Put the plate back into its
packaging and return it to the freezer.
High temperature
Wait at least 5 seconds before taking out the Fluorescence test plate from the 384-well block. The
Fluorescence test plate is heated to a temperature of 100°C and may cause a burn.
6)
The
‘Background
Calibration’
part
will
turn
green(completed) when the Background calibration
is completed.