background image

 

 

 
 

 

 
 
 
 
 
 
   
 
 
 
 
 

 

 
 

 
 
 

 

 

 
 
 
 
 
 
 
 
 
 

 
 
 
 

 

Exicycler

  384

 

 

Real-Time Quantitative Thermal Block 

35

 

15

 

MT Promedt Consulting GmbH Altenhofstr. 80 
D-66386 St. Ingbert. Germany. Tel +49 6894-58 10 20

 

A-2061 

Содержание Exicycler 384

Страница 1: ...Exicycler 384 Real Time Quantitative Thermal Block 35 15 MT Promedt Consulting GmbH Altenhofstr 80 D 66386 St Ingbert Germany Tel 49 6894 58 10 20 A 2061...

Страница 2: ...er 384 Real Time Quantitative Thermal Block User s Guide Bioneer Corporation 8 11 Munpyeongseo ro Daedeok gu Daejeon 34302 Republic of Korea Tel 82 42 930 8777 Fax 82 42 930 8688 Email sales bioneer c...

Страница 3: ...y of the purchaser to determine the suitability of the product for its particular use Liability is conditional upon the customer providing full details of the problem to Bioneer within 30 days Tradema...

Страница 4: ...Exicycler 384...

Страница 5: ...m Views 15 Installation and System Operation 16 Overview 17 Site Requirements 17 Required Materials 17 Unpacking Exicycler 384 17 Setting Up the Computer 18 Connecting the Computer and Exicycler 384 1...

Страница 6: ...1 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Replacing Fuses 63 Returning Exicycler 384 for Service 64 Troubleshooting 65 Ordering Information and Warranty 70...

Страница 7: ...2 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Before Starting...

Страница 8: ...384 please visit us online at www bioneer com How to Reach Customer Support To obtain prompt customer support please call us at 82 42 930 8777 You can also obtain technical support through www bioneer...

Страница 9: ...temp ratures pendant plusieurs heures Risque de condensation pouvant entrainer des fonctionnements d fectueux de l quipement English Be aware of the door Please pay attention to the door when it is mo...

Страница 10: ...t may be displayed on Exicycler 384 Indicates On position of the main power switch Indicates Off position of the main power switch Indicates a standby switch by which the instrument is switched on to...

Страница 11: ...6 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Safety Instructions...

Страница 12: ...icycler 384 and the wall for proper ventilation 5 Do not install Exicycler 384 in a dusty environment to prevent false operation or technical damage 6 Keep Exicycler 384 away from heat sources 7 Exicy...

Страница 13: ...turn off the computer and unplug the power cable before connecting it to Exicycler 384 If the computer is turned on while connecting the communication port connector either in the computer or Exicycl...

Страница 14: ...Time PCR run is done Wait until the cooling fan has stopped running completely Cooling fan remains running for about 2 minutes to cool down the lamp after each Real Time PCR run 6 Do not cover Exicyc...

Страница 15: ...10 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Introduction...

Страница 16: ...ing technique based on polarization of light enhances detection sensitivity for robust and reliable results Patented KR10 1089045 US8427643 polarizing optical apparatus mitigates the common problem of...

Страница 17: ...12 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 System Components and Specifications...

Страница 18: ...ystem component listed below System Components Component Lists Check Exicycler 384 1 USB 2 0 high speed cable 1 Power cord 1 1 Software 1 CD User Guide 1 Optical tape sealing applicator 1 1 Provided a...

Страница 19: ...urce Short arc lamp 120W Sensor 16 bit 2D CCD Excitation Filter Emission Filter 5 Set Continuously turning on and off the lamp reduces the lamp life span The lamp generates heat when it starts Make su...

Страница 20: ...r com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 irritation Read the MSDS before use and follow the instructions if swallowed or inhaled Wear appropriate protective eyewear clothing and gloves...

Страница 21: ...16 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Installation and System Operation...

Страница 22: ...om water Required Materials Scissors pocket knife or box cutter Unpacking Exicycler 384 Save the packing materials and box in case you need to ship the instrument to Bioneer for service 1 To unpack Ex...

Страница 23: ...SB data transfer 2 It is recommended to install an anti virus software in order for Exicycler 384 to operate safely A firewall should also be setup to prevent unwanted information coming in from exter...

Страница 24: ...ver To install Bioneer Package and Exicycler384 softwares for an Exicycler 384 device under Windows 7 follow the instructions below If a previous version of software has been installed on your compute...

Страница 25: ...all program features Custom installation is recommended only for advanced users Click Next 5 InstallWizard is ready for installation If you want to review or modify your settings click Back Click Inst...

Страница 26: ...allation will begin 9 R program will be installed automatically Please wait until installation is complete 10 Click OK for installing R D COM interface 11 R icon will be shown on the Desktop when inst...

Страница 27: ...rk 1 1 Package run dotnetfx_eng exe file of the C Bioneer_Package Net11 folder 2 Click Yes for installing NET Framework 1 1 Package 3 A setup window asking about the license agreement is displayed 4 S...

Страница 28: ...23 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 6 Installation of NET Framework 1 1 is complete Click OK to finish installation...

Страница 29: ...provided CD you can see setup exe file and double click InstallSheild Wizard to install required items automatically 2 Click Next and go to the next step 3 Click Install to begin installation If you...

Страница 30: ...top these icons are created when installation is completed successfully 7 In the C Exicycler384 Run folder folders and files that are shown on the image below are created Copy Calibration folder from...

Страница 31: ...ld connect Exicycler 384 system to a computer using provided USB cable 12 Turn on Exicycler 384 The status LED will be blue when the power is supplied properly 13 Press the power button for a second t...

Страница 32: ...27 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 16 Now you are ready to operate Exicycler 384 system...

Страница 33: ...this path following message will be displayed 2 Click OK to change the path 3 You should assign the path of the Calibration folder to be copied automatically Click Open and you will see the following...

Страница 34: ...hat the device is not turned on or the appropriate device driver has not yet been installed on the computer If you see this message despite the instrument is turned on then you should re install the d...

Страница 35: ...installation was performed previously Bioneer Package may not be installed correctly Go to Control Panel Programs Add Remove and remove 4 programs from the computer Exicycler 384 V1 1 0 Bioneer packa...

Страница 36: ...sis is complete Exicycler 384 is now ready to operate The status LED will turn red when the self diagnosis fails or excessive physical shock is applied from outside e g Make sure that there are no obj...

Страница 37: ...uble click Run Exicycler384 icon on the desktop again IMPORTANT The following error message will appear when the calibration has not been performed for Exicycler 384 or the calibration data is not pre...

Страница 38: ...options in Probe List or add your own Ensure you specify an appropriate probe option for accurate data analysis 5 Probe Information window will appear Select one or click Add to add additional probe A...

Страница 39: ...ave or open a file b User Select or create the user name for personal account management c Protocol Information Displays a cycling protocol in detail d Temperature Profile Displays a temperate curve o...

Страница 40: ...ific amplification products when SYBR Green is used for Real Time PCR e Gradient is used to evaluate an optimum annealing temperature condition for amplification f Store keeps the 384 well thermal blo...

Страница 41: ...cycling is complete The heated lid will start cooling down automatically Exicycler 384 will keep the main door closed and will maintain thermal block at the set temperature until you stop Exicycler 38...

Страница 42: ...the To field i e 94 a temperature interval in the Between field i e 1 and a hold time in the Hold Time field i e 1 16 Click Append 17 Click Incubate to add finish step i e 25 C 1 min and then click A...

Страница 43: ...temperature in the From field ending temperature in the To field Press Gradient Calculator button below the time setting window and Gradient Calculator window will appear It will show predicted temper...

Страница 44: ...information for each sample c Probe Name Flu Dye Quencher Type Concentration specifies a name a fluorescence dye a quencher dye a type and a concentration for each sample d Protocol Plate tab switches...

Страница 45: ...Select appropriate probes and then click OK a Probe Name displays probe name assigned by the user b Fluorescent Dye displays name of a fluorescence dye assigned by the user c Quencher displays name of...

Страница 46: ...Type drop down list Select the check box of Probe Name and then click Assign a Sample represents unknown samples b STD stands for a standard sample of a known concentration c NTC stands for No Templa...

Страница 47: ...rough E8 and enter a sample name in the Sample Name field Click Sample Name to save sample name 28 Verify information on each well in the Name Count Probe window 29 Enter a plate name in the Plate fie...

Страница 48: ...saved master file by clicking button from Quick Start menu The master file includes both protocol and plate file Power off Standby 32 Ensure the status LED in front of the Exicycler 384 is blinking gr...

Страница 49: ...ing three message boxes will appear in the following order a This message will appear when closing the door of the Exicycler 384 b This message will appear when the lamp condition is being checked c T...

Страница 50: ...f the pop up window remains more than 5 minutes after running Exicycler 384 or if the temperature profile does not display anything in the main window turn off Exicycler 384 and turn it on again Start...

Страница 51: ...l 384 well plate You must use a plate when performing Real Time Quantitative PCR in Exicycler 384 A standard plate can be also used but only for a conventional PCR without fluorescent dyes Optical sea...

Страница 52: ...47 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Performance Maintenance...

Страница 53: ...POWER button 2 Go to C Exicycler384 Run and double click ExiCfg exe to start the program 3 Go to Tools Self Test NOTE USB Communication is NOT initialized message box will pop up if Exicycler 384 is t...

Страница 54: ...ll of the self diagnosis tests pass completely without any problems NOTE Self diagnosis is complete with two short beeps 6 When any of the self diagnosis tests fail NG signs will appear in red NOTE An...

Страница 55: ...he Fluorescence test kit is available for reuse up to 3 times For more information about the Fluorescence test kit please see the following Before Starting Calibration with the Fluorescence test kit C...

Страница 56: ...e test plate in a rack and centrifuge for 5 minutes NOTE It is important to centrifuge the Fluorescence test kit prior to use This will spin down remains of fluorescence dye on the side of the well an...

Страница 57: ...nging a light source lamp A periodical calibration of every 6 months helps to maintain the optimum condition of Exicycler 384 Use the ExiCfg program to calibrate Exicycler 384 The ExiCfg calibrates as...

Страница 58: ...84 and turn it on again after 5 seconds In case of an abnormal shutdown like this wait about 5 minutes then click Lamp warming to turn on the lamp again It takes approximately 10 minutes to reach the...

Страница 59: ...again to close the door 7 Click Mask Calibration Check button 8 Click Start button in the Mask Calibration window NOTE The Calibration time is approximately 10 minutes depending on the condition of Ex...

Страница 60: ...e plate back into its packaging and return it to the freezer High temperature Wait at least 5 seconds before taking out the Fluorescence test plate from the 384 well block The Fluorescence test plate...

Страница 61: ...k Calibration 2 Place the Background Fluorescence test plate on the block IMPORTANT It is important to load the Fluorescence test plate in the right position A misplaced plate may cause damage or tech...

Страница 62: ...ccessfully complete Put the plate back into its packaging and return it to the freezer High temperature Wait at least 5 seconds before taking out the Fluorescence test plate from the 384 well block Th...

Страница 63: ...e when performing the Background Calibration followed by the Mask Calibration 2 Click Multi Channel Calibration button 3 Select fluorescence dyes in the Multi Channel Calibration window for the Calibr...

Страница 64: ...e plate named X i e FAM on the bath message appears 6 Place the Multi Channel Fluorescence Test plate on the block IMPORTANT It is important to load the Fluorescence Test plate in the right position A...

Страница 65: ...e test plate from the 384 well block The Fluorescence test plate is heated to a temperature of 100 C and may cause a burn 9 Prepare the Fluorescence Test plate with a dye X i e SYBR Green when Attach...

Страница 66: ...t from the top menu bar to close ExiCfg program IMPORTANT Wait at least 5 minutes to switch off main power of Exicycler 384 The cooling fan will continue to run to cool down the lamp Shutting down Exi...

Страница 67: ...h block and set it aside 3 Use a cotton swab soaked in pure isopropyl alcohol or DNA Zap to clean the wells thoroughly Make certain that the isopropyl alcohol or DNA Zap has evaporated completely befo...

Страница 68: ...igh voltage power supply To avoid electrical shock disconnect the power supply to the instrument unplug the power cord and wait at least 1 minute before working on the instrument 2 Insert a small scre...

Страница 69: ...to Bioneer for service 2 Complete certification for instrument decontamination 3 Fax or e mail exicycler support bioneer com the service request form to the customer center 4 Pack the instrument in th...

Страница 70: ...65 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Troubleshooting...

Страница 71: ...f the same error keeps occurring No communication detected between the computer and Exicycler 384 1 Make sure that the USB cable is firmly connected to the computer 2 Ensure you use the USB cable prov...

Страница 72: ...ng lid works or if the sample tubes or plates are completely sealed with the sealing tape 2 Ensure the heating lid option is set up correctly The Hottop Check Box must be selected and correct temperat...

Страница 73: ...ating with each other Check if the USB cable between the computer and Exicycler 384 is firmly connected 13 You must insert New Dye Name will appear if a new name is not assigned to a custom dye 14 You...

Страница 74: ...ta ID do not match 14 Please insert any protocol before inserting Goto Protocol will appear if the Goto step is inserted in the first line of a protocol file 15 Please insert Data name will appear if...

Страница 75: ...70 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Ordering Information and Warranty...

Страница 76: ...Real Time Quantitative Thermal Block Plastic consumables 3111 41 Adhesive Optical Sealing Film 100 sheets Premix Reagent K 6253 AccuPower 2X Greenstar qPCR Master Mix 100Rxn 50 L reaction K 6254 AccuP...

Страница 77: ...for five years from the initial instrument release date If a defect arises after the limited warranty period shipping and handling charges may apply to any repairs or exchanges of the product underta...

Страница 78: ...bioneer com www bioneer com 155FilbertSt Suite216Oakland CA94607 USA 1 877 264 4300 Toll free 1 510 865 0350 order usa bioneer com us bioneer com KoreaBioParkBLDG B 702 700Daewangpangyo ro Bundang gu...