Figure 10
Figure 11
11
Femoral Preparation
Distal Femoral Resection (cont.)
• Assemble the IM rod (32-485030) and adjustable distal
cut guide (32-487000) by inserting the IM rod through
the central hole of the adjustable distal cut guide.
• Slowly introduce the IM rod to the femoral canal to
depressurize the canal.
• Slide the adjustable distal cut guide until it rests flush
with the distal femur.
• Attach the distal cut block (32-487002) to the adjustable
distal adaptor by sliding the magnetized distal cut
block into the adjustable distal adaptor (32-487010).
• Attach the adjustable distal adaptor and distal cut
block to the adjustable distal cut guide by sliding the
two legs on the adjustable distal adaptor through
the anterior holes of the adjustable distal cut guide
(Figure 10).
• Continue sliding the adjustable distal adaptor until the
block is sitting against the anterior cortex.
• Pin the distal cut block into place using 1/8" quick
release drill pins (32-467619) in the center pinholes of
the block (Figure 11).
Note: To confirm the valgus angle, the alignment handle
can be inserted into the adjustable distal adaptor and a 1/4"
alignment rod (32-486135) can be inserted and extended to
the center of the femoral head.