Overview
Intact ACL and PCL
Intact, Functioning PCL
Partially or Non-functional PCL
Femoral Component
XP Femur
XP Femur
XP Femur
Tibial Component
XP Tibia
XP or CR Tibia
XP or CR Tibia
Bearing
XP- XP Bearings
XP- XP Bearings
XP- AS Bearings
If additional implant constraint is desired, the Vanguard PS or PSC system can be utilized.
4
VANGUARD
X
P
Total Knee System
The Vanguard XP Total Knee System consists of:
• An ACL/PCL retaining femoral component
• Two tibial trays: XP and CR
• Two bearing designs: XP-XP and XP-AS
The components can be used in different combinations, depending on the patient’s soft tissue status (see table below).
Note:
•
Femoral Pegs: The Vanguard XP femoral component has two modular distal pegs. To help address minor
bone deficiencies, the pegs can be removed and distal femoral augments attached using an augment
bolt or an augment peg.
•
Femoral Augments: The use of femoral augments does not necessarily change the recommended
implant combinations (table above) – the appropriate implant combination should be chosen based on
the intact soft tissues.
Description
Biomet manufactures a variety of knee joint replacement prostheses intended for application with or without bone
cement. Knee joint replacement components include femoral, tibial, and patellar components. Components are available
in a variety of designs and size ranges intended for both primary and revision applications. Specialty components are
available including femoral stems, femoral screws, femoral augments, tibial stems, tibial screws, tibial augments, tibial
cement plugs and modular pegs.
• The Vanguard XP Knee System offers the flexibility to change from an ACL/PCL retaining (XP) to a PCL retaining (CR),
or non-functional PCL (AS) within a single system.
• The transition between each constraint level can be made with ease, allowing the physician to evaluate soft tissue and
bone deficiencies intraoperatively without making a preoperative commitment to the level of constraint.
• The Vanguard XP femoral components (PN 195905-195952) may also be used with the Vanguard standard line
(i.e., CR, CR-L and AS bearings) E1 Polyethylene and UHMWPE (Arcom) bearing components (when used
with Vanguard tibial trays).
Note: Improper surgical technique may result in damaging/impinging the ACL, undercutting/over-cutting/
over-cutting/fracture of the tibial bone island, and/or difficulty in component fit/sizing with patient
anatomy soft tissues.