background image

TABLE OF CONTENTS

1. GENERAL ------------------------------------------------------------ 1

2. TECHNICAL SPECIFICATIONS----------------------------------- 2

3. PACKING CONTENT ----------------------------------------------- 3

4. INSTALLATION ---------------------------------------------------- 4

5. CONTROL PANEL --------------------------------------------------- 5

6

6.1 POWER ON

------------------------------------------------------

6

6.3 PREPARING THE MATERIAL FOR STERILIZATION -----

6

6.4 SELECTING THE STERILIZATION PROGRAM -------------- 7

6.5 RUNNING THE STERILIZATION PROGRAM ----------------- 7

7. DRAINING THE WATER TANKS ---------------------------------- 9

8. ADVANCED SETTING ---------------------------------------------10

9. PRINTER (OPTIONAL) --------------------------------------------11

10. MAINTENANCE -------------------------------------------------- 12

11. TRANSPORT AND STORAGE ----------------------------------- 15

12. ERROR CODES-----------------------------------------------------16

13. SERVICE AND CONTACT---------------------------------------- 17

. OPERATION --------------------------------------------------------- 6

6.2 FILLING THE DISTILLED WATER -----------------------------

6

1. WATER PROPERTIES / CHARACTERISTICS -------------- 18

2. DIAGRAMS OF THE STERILIZATION PROGRAMS -----

19

3. ELECTRICAL DRAWING ----------------------------------------- 23

4. HYDRAULIC DRAWING  -----------------------------------------   24

APPENDIX 3

2 3

APPENDIX

Instructions manual

Instructions manual

ELECTRICAL DRAWING

Tp1: Steam generator temperature sensor
TP2: Inner-Chamber temperature sensor
TP3: Temperature sensor of chamber wall
V1: Air release valve
V4: Water release valve

¡

220V

T12A

T12A

¡

21V

¡

9V

V4

V1

FAN

AC220V
OUTPUT

WATER
PUMP

STEAM
HEATER

CHAMBER
HEATER

THERMAL
PROTECTOR

THERMAL
PROTECTOR

PRESSURE SENSOR

KEYBOARD
DATA LINE

TP 1

TP 3

TP 2

PRINTER
OUTPUT

USED WATER TANK
MAX. LEVEL

TANK MIN. LEVEL

TANK MAX. LEVEL

DOOR SWITCH

PUBLIC

110

20

20

110

Содержание BioClave

Страница 1: ...Operations Manual V 1 0...

Страница 2: ...you for choosing the Biomega BioClave steam sterilizer Your steam sterilizer has been CE certified and designed with durability reliability and safety in mind It is your responsibility to install thi...

Страница 3: ...LED WATER 6 1 WATER PROPERTIES CHARACTERISTICS 18 2 DIAGRAMS OF THE STERILIZATION PROGRAMS 19 3 ELECTRICAL DRAWING 23 4 HYDRAULIC DRAWING 24 APPENDIX 3 23 APPENDIX Instructions manual Instructions man...

Страница 4: ...ols It operates automatically with 134 C and 121 C sterilization temperatures This sterilizer is in compliance with the European Directive 93 42 CEE and it has been produced in accordance with the EN...

Страница 5: ...0Hz 3 Nominal power 1600W 120V 1250W 220V 4 Sterilization Temperature 121 C 134 C 5 Main Fuses T20A 250V 6 Capacity of the distilled water tank Approx 2 5L water at level MAX 7 OperatingEnvironment 5...

Страница 6: ...tray Tray rack Tray handling tool Door adjustment tool Draining hose Instructions manual Spare fuse power Spare fuse for valve 3A Door seal 3 1 1 1 2 1 2 2 2 1 1 16L Steam sterilizer 1 Pressure psi U...

Страница 7: ...Wrapped liquid 0 50 0 50 Ensure that the sterilizer is installed with 2 5in 10cm ventilation space on all sides of the sterilizer and 5 in 20cm on top side The clearance required to open the door is...

Страница 8: ...le To cancel or stop the cycle press and hold this button for 3 sec Time Display State State description Ld HE Time in Min PL Ed E1 E9 Do preheating loading Door state vacuuming heating sterilization...

Страница 9: ...3 Sample Preperation For the most effective sterilization and to preserve the sample please follow below Arrange the samples of different material on different trays or with at least 2in of space betw...

Страница 10: ...pump or the door seal Check the air release valve Make sure you have turned the door handle to the max position or check the door switch Check the water pump and the air release valve Check steam gene...

Страница 11: ...ton is pressed again do After a cycle is completed the time display will show Ed Once the pressure has reached 0 PSI on the display the door can be opened and samples can be removed 6 8 Sterilization...

Страница 12: ...ank The right valve is used for draining the clean water tank Always be sure that the hose is draining into a vessel with at least 2L capacity 10 5 1 Fully open the door 10 5 2 Remove the door seal ca...

Страница 13: ...the program there is only one option 2 This is to adjust the drying time TEMP START DRYING Caution Always ensure that the sample being sterilized can withstand the sterilization time that is being pr...

Страница 14: ...back of the sterilizer 9 2 Connect the printer power cord to the socket with the proper electrical specifications 9 3 Power on the sterilizer and press the sel button 9 4 After the sterilization cycle...

Отзывы: